pH Sensors and Electrodes “Maintaining your Yokogawa pH Monitoring System”
●When a quality pH sensor system is undamaged, cleaned and properly calibrated it will provide a measurement that is accurate and reliable. Even if an electrode is in a process that does not cause coating, plugging, abrasion or other problems, it still must be calibrated periodically to correct for aging of the sensors and the non-recoverable changes to the electrodes that take place.
●These effects of aging usually happen slowly therefore, calibration should not be necessary more frequently than about once a month. If more frequent calibration is called for, it is usually because (1) the process is aggressive; (2) the electrode cleaning process used was not effective; (3) the routine calibration was not properly executed; (4) the pH readings are temperature dependent or (5) the wrong electrodes have been selected.
●Users desire an accurate, reliable measurement with a reasonable electrode life while minimizing the frequency or complexity of maintenance that is needed. However, ensuring the electrodes are clean and calibrated will always involve some amount of maintenance.
●The measurement results obtained from dirty or faulty electrodes can be anything from slow response, to a completely erroneous reading.
●Therefore, the accuracy of the pH measurement is directly dependant upon the frequency of maintenance, and the maintenance frequency (and also the method used) is directly dependent upon the importance and/or aggressiveness of the process.
|
|
Application note & Design Guide |
|
|
|
Please see the document for details |
|
|
|
|
|
|
|
English Chinese Chinese and English Japanese |
|
April 1 2009 |
|
|
|
TNA0918 |
|
137 KB |
- +1 Like
- Add to Favorites
Recommend
- Transformer Characteristic Parameters
- Working Principle of Transformer
- Principles of Electronic Transformers
- Exchange Voltage Conversion Techniques “Authentic“ Electronic Transformers
- Overview of the Electronic Transformer Industry
- IoT-based Water Quality Monitoring, How it Works? How to Implement?
- Calculation Method for Melt Current of Fuse Fuses
- Kinghelm‘s PH2.0 Terminal KH-PH-PT2.0 Sets New Standards Due to Its Outstanding Performance and Excellent Stability
All reproduced articles on this site are for the purpose of conveying more information and clearly indicate the source. If media or individuals who do not want to be reproduced can contact us, which will be deleted.