Synchronising Xsens Systems with TMSI Porti Step-by-step manual

2020-12-17

The steps described below show how to make it possible for Xsens Awinda or Sync Station to send a trigger pulse to TMSI's EMG system Porti. Using Xsens systems, a trigger pulse is recognized in the digital channel of TMSI's Polybench software (the software reading data from the Porti, Mobi, Reva and Mobita systems). The tested synchronization was based on the Porti system.
1.Perform the calibration and get both systems fully set.
2.Set the specifications for synchronization in both systems and connect sync cable(s)
3.Start the recording on the slave software. The slave software will wait for a trigger from the Master software to start recording.
4.Start the recording on the Master software. Then, both systems will start recording at the same time.
5.Stop recording on the master software. Then both systems will stop recording at the same time.
Depending on the type of synchronization performed different specifications are required for both systems. A detailed explanation of the different characteristics according to the desired synchronization are described below.
This document is written to synchronize the XSens Awinda system. However, most of the steps performed in the MVN software are the same for the MVN Link system. You can see the differences in the setup of the system in the last chapter of this document “Using MVN Link to synchronize”.

XSENS

More

More

User's Guide

More

More

Please see the document for details

More

More

English Chinese Chinese and English Japanese

16-01-2020

645 KB

- The full preview is over. If you want to read the whole 8 page document,please Sign in/Register -
  • +1 Like
  • Add to Favorites

Recommend

All reproduced articles on this site are for the purpose of conveying more information and clearly indicate the source. If media or individuals who do not want to be reproduced can contact us, which will be deleted.

Contact Us

Email: