TBS1000B and TBS1000B-EDU Series Oscilloscopes Specifications and Performance Verification Technical Reference

2022-08-04

●Specifications
■These specifications apply to all TBS1000B series oscilloscopes. To verify that an oscilloscope meets specifications, it must first meet the following conditions:
▲The oscilloscope must have been operating continuously for twenty minutes within the specified operating temperature.
▲You must perform the Do Self Cal operation, accessible through the Utility menu, if the operating temperature has changed by more than 5 °C (9 °F) since the last time the Do Self Cal operation was performed.
▲The oscilloscope must be within the factory calibration interval of one year.
■Specifications are provided in the following tables. All specifications are guaranteed unless noted "typical." Specifications that are marked with the symbol are checked in the Performance Verification section. (See page 18,Performance Verification.)

TEKTRONIX

TBS1000BTBS1000B-EDU077-1025-01TBS1032BTBS1202BTBS1202B-EDUTBS1152BTBS1152B-EDUTBS1102BTBS1102B-EDUTBS1072BTBS1072B-EDUTBS1052BTBS1052B-EDU5500A-SC5500A012-0482-XX012-0208-XX011-0049-XX103-0090-XX103-0030-XX015-0565-XX103-045-XX103-0058-XX077-1025-xx.077-1025-xx.077102501

More

Part#

Oscilloscopes

More

More

User's Guide

More

More

Please see the document for details

More

More

English Chinese Chinese and English Japanese

2017/12/22

Revision B

2.2 MB

- The full preview is over. If you want to read the whole 49 page document,please Sign in/Register -
  • +1 Like
  • Add to Favorites

Recommend

All reproduced articles on this site are for the purpose of conveying more information and clearly indicate the source. If media or individuals who do not want to be reproduced can contact us, which will be deleted.

Contact Us

Email: