HUMIDITY / TEMPERATURE / DEW POINT / FROST POINT CALIBRATION

2022-11-11
■ The procedure of dew point and frost point calibrations are equal with the procedure of humidity calibrations!
●The EE23 transmitter series can be calibrated in two ways:
◆1-point humidity / temperature calibration: quick and simple calibration on a defined humidity / temperature point(working point).
◆2-point humidity / temperature calibration: simple calibration for accurate measuring results over the whole humidity / temperature working range.
●To reach a temperature balance it is recommended to keep the transmitter and the reference chamber (e.g. HUMOR 20,...) for minimum 4 hours in the same room.
●During calibration procedure it is important to keep the temperature constant!
●For calibration the humidity sensor probe must be stabilised at least 30 minutes in the reference chamber.
●Replace a used dirty filter cap before calibration!