MOX Series and INQOX Medical Oxygen Sensors Operating Manual

2022-06-08
The sensor should be replaced only by a professional user. Before insertion into the device, check the sensor for mechanical damages and for crystallization of salts on the housing-do not use a damaged sensor or a sensor with crystallization of salts outside. Follow the instructions for use of the oxygen analyzer for replacement of the sensor. Verify that the sensor can be properly attached to the mechanical and electrical connections of the oxygen analyzer. Calibrate the analyzer according to the instructions in the analyzer’s operation manual and verify proper gas readings. Oxygen analyzer readings in room air will typically be between 19% and 23% when calibrated in 100% oxygen or another calibration gas level required in accordance with the analyzer’s instructions. The sensor should be calibrated in regular intervals (see instructions for use of the analyzer). If calibration problems or unstable signals occur, the sensor must be replaced.