Models 2634B, 2635B, and 2636B System SourceMeter Instrument Specifications
★This document contains specifications and supplemental information for the Models 2634B, 2635B, and 2636B System SourceMeter™instrument. Specifications are the standards against which the 2634B, 2635B, and 2636B are tested. Upon leaving the factory,the 2634B, 2635B, and 2636B meet these specifications. Supplemental and typical values are nonwarranted, apply at 23 °C, and are provided solely as useful information.
★Source and measurement accuracies are specified at the 2634B, 2635B, and 2636B terminals under these conditions:
●18°C to 28°C, < 70 percent relative humidity
●After a two-hour warm-upperiod
●Speed normal (1 NPLC)
●A/D autozero enabled
●Remote sense operation or properly zeroed local operation
●Calibration period: One year