Models 2601B, 2602B, and 2604B System SourceMeter® Instrument Specifications

2022-05-30
●SPECIFICATION CONDITIONS
■This document contains specifications and supplemental information for the Models 2601B, 2602B, and 2604B System SourceMeter® instrument. Specifications are the standards against which the Models 2601B, 2602B, and 2604B are tested. Upon leaving the factory, the Models 2601B, 2602B, and 2604B meet these specifications. Supplemental and typical values are nonwarranted, apply at 23 °C, and are provided solely as useful information.
■Source and measurement accuracies are specified at the Models 2601B, 2602B, and 2604B terminals under these conditions:
▲18°Cto 28°C, < 70percent relative humidity
▲After a two-hour warm-upperiod
▲Speed normal (1 NPLC)
▲A/D autozero enabled
▲Remote sense operation or properly zeroed local operation
▲Calibration period: One year