Model 2601B-PULSE-CA1 Cable Kit Instructions
Two of these cables are jumpered to reduce loop inductance and are intended for making the FORCE HI and LO connections for a 4-wire measurement. The jumpered end is intended for connection to the device under test (DUT).
The other two cables are intended for making the SENSE HI and LO connections.
This cable kit is intended for use with the Model 2601B-PULSE System SourceMeter®instrument with the 2601B-P-INT Interlock and Cable Connector Box installed.
●Characteristics
■SENSE cables
▲Rated voltage and current: 40 V maximum, 11 A
▲Operating environment: −40 °C to +85 °C
▲Characteristic impedance: 50 Ω
▲DC resistance: 70 mΩ
▲Cable diameter: 2.85 mm (0.112 in.)
■FORCE cables
▲Rated voltage and current: 40 V maximum, 11 A
▲Operating environment: −40 °C to +85 °C
▲Characteristic impedance: 50 Ω
▲Loop DC resistance*: 150 mΩ
▲Loop inductance at 100 kHz*: 700 nH
▲Cable diameter: 2.85 mm (0.112 in.)
▲Jumper length: 152 mm (6.0 in.)
|
|