Performance Test of Renesas OB1203 Using OB1203SD-RL-EVK

2021-02-26
● Objective:
The objective of this test is to qualify the OB1203 for use in pulse oximeters across the range of human characteristics including skin color and tissue blood perfusion.
● Abstract:
A method was developed to test performance on dark skin and low perfusion using a simulator. The OB1203 Biosensor was tested.
We conclude that the OB1203-RL-EVK hardware and demo algorithm are capable of measuring accurate saturations on the most difficult subjects normally encountered in clinical testing, those with dark skin and low perfusion. The perfusion level we were able to achieve was 0.3% which would be sufficient for accurate monitoring for dark skinned and very purely perfused human subjects.