MSO4000 and DPO4000 Series Digital Phosphor Oscilloscopes Specifications and Performance Verification
●Specifications
■This chapter contains specifications for the DPO4000 and the MSO4000 series oscilloscopes. All specifications are guaranteed unless noted as "typical."Typical specifications are provided for your convenience but are not guaranteed.Specifications that are marked with the symbol are checked in Performance Verification.
■All specifications apply to all DPO4000 and MSO4000 models unless noted otherwise. To meet specifications, two conditions must first be met:
▲The oscilloscope must have been operating continuously for twenty minutes within the operating temperature range specified.
▲You must perform the Signal Path Compensation (SPC) operation described in the Tektronix 4000 Series Digital Phosphor Oscilloscopes User Manual prior to evaluating specifications. If the operating temperature changes by more than 10 °C (18 °F), you must perform the SPC operation again.
|
|