OM-CP-PRTEMP1000IS Intrinsically Safe Pressure and Temperature Data Logger User's Guide
■Getting Started
▲To access the COM Port for the interface cable, unscrew the key-ring end cap. Screw the end cap back onto the data logger until the o-ring cannot be seen, before deploying the data logger.
■Intrinsic Safety Approval
▲The OM-CP-PRTemp1000IS has been certified by FM Approvals as Intrinsically Safe (IS) for use in Class I, Division 1, groups A, B, C, D, and Non-incendive (NI) for use in Class I, Division 2, groups A, B, C, D Hazardous (Classified) Locations. The rating listed in the Factory Mutual approval guide is as follows: OM-CP-PRTemp1000IS. Pressure & Temperature Recorder. IS / I / 1 / ABCD T4A Ta=80 °C; NI / I / 2 / ABCD / T4A Ta=80 °C
■Pressure Sensor
▲To use the pressure sensor for gauge measurements, screw the ¼" male NPT fitting into the pipe to be measured with a 9/16"wrench. The NPT fitting can be used for absolute and submersible applications, but it is preferable to use the submersible fitting.
■O-Rings
▲O-ring maintenance is a key factor when properly caring for the
OM-CP-PRTEMP1000IS. The O-rings ensure a tight seal and prevent liquid from entering the inside of the device.
|
|