OM-CP-EVENT101A EVENT DATA LOGGER INSTRUCTION SHEET
■The OM-CP-EVENT101A data logger measure a digital signal. It records events (i.e. on/off, open/closed) that occur at one second intervals or slower. At the specified reading rate, the device will take a reading and write whether an event is occurring, to memory. If there is a change in voltage above 2.7V's or below 0.4V's, the device will register that an event occurred. It is important to choose a reading rate that is faster than the event could occur, to ensure that it will be captured by the data logger. The OM-CP-EVENT101A can accept up to 30VDC continuous.
|
|