5 Series MSO MSO54, MSO56, MSO58, MSO58LP Specifications and Performance Verification
●To meet specifications, these conditions must first be met:
■The instrument must have been calibrated in an ambient temperature between 18 °C and 28 °C (64 °F and 82 °F).
■The instrument must be operating within the environmental limits. (See Environmental specifications on page 29.)
■The instrument must be powered from a source that meets the specifications. (See Power supply system on page 29.)
■The instrument must have been operating continuously for at least 20 minutes within the specified operating temperature range.
■You must perform the Signal Path Compensation procedure after the warm up period. See the online help for instructions on how to perform signal path compensation. If the ambient temperature changes more than 5 °C (9 °F) , repeat the procedure.
|
|