ULTRARAE AND ULTRARAE 3000 OPERATION TIPS
●When inserting a RAE-Sep tube into the inlet adapter, push the tube into the rubber gasket until you feel that the tube is properly sealed. Next screw on the tube holder cap to seal the other end of the RAE-Sep tube. Then simultaneously twist and push down on the tube cap nut to ensure that the rubber adapters inside the tube holder form good seals against both ends of the RAE-Sep tube.
●Before performing a compound-specific measurement for Benzene or Butadiene using a RAE-Sep separation tube, make sure the UltraRAE 3000 is in Tube Mode and that the appropriate tube type is selected. The UltraRAE 3000 only acts as a compound-specific measurement device when it is equipped with a 9.8eV lamp. Therefore, make sure the UltraRAE 3000 is equipped with a 9.8eV lamp and manually set to use it whenever it is to be used for compound-specific sampling such as for benzene.
|
|