GL90B PREAMPLIFIER WIRING GREAT LAKES MODEL 672

2022-05-12
■NOTES :
●GENERAL
◆OUTPUT TO THE INSTRUMENT FROM THE PREAMP IS UNITY GAIN.
◆WHEN A COMBINATION ELECTRODE IS USED, REFERENCE CONNECTION IS MADE THROUGH THE BNC CONNECTOR.
●TEMPERATURE COMPENSATION
◆CONNECT A 3KΩ @25°C BALCO TC ELEMENT TO TERMINALS 4 AND 5 OF THE PREAMP, AS SHOWN. THEN PUT A 301Ω RESISTOR ACROSS 4 AND 5(BLACK AND YELLOW) IN THE INSTRUMENT.
●EARTH GROUNDING WARNING
◆DON'T LET ANY OF THE LEADS OR THE BNC CONNECTOR TOUCH METAL PIPING, CONDUIT, OR OTHER EARTH GROUNDS. THIS WILL CREATE A"GROUND LOOP" AND CAUSE VERY UNSTABLE OR OFF-SCALE READINGS. TO CHECK FOR A GROUND LOOP, DISCONNECT THE BNC AND PUT AN OHMMETER BETWEEN TERMINAL 5 OF THE PREAMP AND A GOOD EARTH GROUND. YOU SHOULD READ INFINITY. ANY OHMS READING AT ALL INDICATES A SHORT TO GROUND WHICH MUST BE FOUND AND ELIMINATED.