HHM-380395 Digital High Voltage Insulation Tester
■Congratulations on your purchase of the Extech 380395 (120V) or 380396 (220V) High Voltage Insulation Tester. This meter provides four Insulation Resistance test ranges plus Continuity, AC/DC Voltage, Polarization Index and Dielectric Absorption Ratio measurements. This meter is shipped fully tested and calibrated and, with proper care and use, will provide years of reliable service.
●Features
■Auto ranging 6000 count LCD display with bargraph
■Selectable LCD backlight level function facilitates working in dimly illuminated areas.
■MAX/MIN, PEAK, Relative, and Data hold functions for DC/AC voltage measurements.
■Designed to the following safety standards:
▲IEC 61010-1 (CAT IV 600V Pollution degree 2)
▲IEC 61010-031 (Requirements for hand-held probes)
■Insulation test range: 0.1MΩ to 60GΩ.
■Insulation test voltages: 500V, 1000V, 1500V, and 5000V.
■AC/DC voltage: 0.5 V to 600 V.
■200mA Continuity.
■Resistance: 0.1 Ω to 6kΩ.
■Auto-discharge function & voltage output warning function.
■Electric charges stored in capacitive circuits are automatically discharged after measurements. Discharge status can be checked using the real time voltage bargraph.
■LIVE circuit warning symbols plus audible warning.
■Live circuit detection prevents insulation test if voltage > 30 V is detected.
■Auto-power off function and battery check.
■Programmable Test Timer measurement function.
■Polarization Index measurement (PI).
■Dielectric Absorption Ratio measurement (DAR).
■Power source: Eight (8) 1.5V 'C' Cell Batteries.
|
|