M7010-V013 SECUTEST SIII+ | . . . Technical Data

2022-04-14
●Test Instrument for DIN EN 60601/60335/60950/61010, DIN VDE 0700/0701-0702 and IEC 62353 (VDE 0751-1).
●The SECUTEST SIII+. . .test instruments can be configured for international utilization. The test socket, user interface language and the desired test regulation can all be configured to this end.
●Universal test instrument for testing the electrical safety of:
■after repair and for periodic testing of electrical devices in accordance with DIN VDE 0701-0702:2008.
■after repair and for periodic testing as well as for technical safety checks of electrical medical devices per IEC 62353 (VDE 0751-1) and/or DIN EN 60601 on the basis of the MPG¹).
■in the production of:
◆electrical measuring, control and laboratory devices per DIN EN 61010.
◆data processing systems per DIN EN 60950.
◆electrical medical devices per DIN EN 60601.
◆household appliances per DIN EN 60335.
●DAkkS Calibration Certificate included.
●Power shutdown as soon as the leakage current exceeds approx. 15 mA provides a maximum of safety for the user.
●High-voltage test for: DIN EN 60950, DIN EN 61 010, DIN EN 60335 and DIN EN 60601.
●Features:
■The device under test can be connected:
◆to the test socket with or without adapter for various types of mains connection.
◆to the connector jacks if the device under test does not have a mains plug.
◆with an adapter for extension cables with or without multiple outlet sockets.
◆connection for BE 2) and FE 3).
◆10 application parts can be connected individually or in groups.
■Mains Plug Polarity Reversal:
◆Mains plug polarity need not be reversed manually. Polarity reversal is accomplished internally during the test sequence (except if 3-phase current adapters are used).
■Automatic Recognition:
◆of mains connection errors and protection class (I or I I ). Measurement is automatically disabled in the event of danger.
■Display:
◆Menus, setting options, measurement results, instructions and error messages, as well as online help and schematic diagrams for test setups, can all be displayed at the backlit, dot matrix LCD.
★MPG = German medical product law;
★BE = Operational earth;
★FE= Functional earth.
■Menu Driven Test Sequences:
◆Fully automatic or manual.
■Selectable Test Current for Protective Conductor Testing (4-pole Measurement):
◆With 200 mA, 10 A or 25 A test current.
■Insulation Test:
◆By means of insulation resistance or equivalent leakage current measurement or high-voltage test.
■Leakage Current Test:
◆With measurement of earth, housing or patient leakage current, patient auxiliary current, protective conductor current, contact current, verification of absence of voltage by means of current measurement or device leakage current.
■Basic Instrument and Expansion Features:
◆The test instrument can be configured with specific features required for the given application (see table on page 6).
■Data Interface for PC, Printer and Barcode.
■Expandable:
◆The SECUTEST SI+ option expands the basic instrument into a unique data logger with memory and alphanumeric keypad for data entry.
◆All required reports can be generated, and data can be analyzed and managed with the help of user-friendly WINDOWS software.

Gossen Metrawatt

M7010-V013M7010M702FGTM5016000R0001Z745UZ853HZ745DZ745NZ745KZ745GZ745LGTZ3409000R0001GTZ3408000R0001Z219CZ864AZ723AGTZ3225000R0001GTZ3226000R0001GTZ3226000R0002GTZ3227000R0001GTZ3219000R0001Z501AGTZ3214000R0003GTZ3214000R0002GTZ3228000R0001Z745WZ750AZ750BZ745TZ745XZ745SZ745AGTY3624065P01GTY3620034P0002Z715AZ700DZ700GZ700F

More

Part#

Test Instruments

More

More

Datasheet

More

More

Please see the document for details

More

More

English Chinese Chinese and English Japanese

18/3.19

3-349-397-03

403 KB

- The full preview is over. If you want to read the whole 10 page document,please Sign in/Register -
  • +1 Like
  • Add to Favorites

Recommend

All reproduced articles on this site are for the purpose of conveying more information and clearly indicate the source. If media or individuals who do not want to be reproduced can contact us, which will be deleted.

Contact Us

Email: