5 Series MSO MSO54, MSO56, MSO58, MSO58LP Specifications and Performance Verification
■This chapter contains specifications for the instrument. All specifications are guaranteed unless noted as "typical." Typical specifications are provided for your convenience but are not guaranteed. Specifications that are marked with the √ symbol are checked in this manual. All specifications apply to all models unless noted otherwise.
■To meet specifications, these conditions must first be met:
▲The instrument must have been calibrated in an ambient temperature between 18 °C and 28 °C (64 °F and 82 °F).
▲The instrument must be operating within the environmental limits. (See Environmental specifications on page 34.)
▲The instrument must be powered from a source that meets the specifications. (See Power supply system on page 34.)
▲The instrument must have been operating continuously for at least 20 minutes within the specified operating temperature range.
▲You must perform the Signal Path Compensatio n procedure after the warmup period. Seethe online help for instructions on how to perform signal path compensation. If the ambient temperature changes more than 5 °C (9 °F), repeat the procedure.