Model 2410 1100 V SourceMeter Instrument

2021-07-21
●Introduction
■Use the procedures in this section to verify that Model 2410 accuracy is within the limits statedin the instrument's one-year accuracy specifications. You can perform these verification procedures:
▲When you first receive the instrument to make sure it was not damaged during shipment.
▲To verify that the unit meets factory specifications.
▲To determine if calibration is required.
▲Following calibration to make sure it was performe d properly.
■WARNING
The information in this section is intended for qualified service personnel only. Do not attempt these procedures unless you are qualiiied to do so. Some of these procedures may expose you to hazardous voltages, vdiich could cause personal injury or death if contacted・ Use standard safety precautions when working with hazardous voltages.
■NOTE
If the insirumem is still under -warranty and iis performance is outside specified limits, contact your Keithley representaiive or the fiiciory to determine the correct course of action.

TEKTRONIX

2410-902-01C2410

More

Part#

SourceMeter Instrument

More

More

User's Guide

More

More

Please see the document for details

More

More

English Chinese Chinese and English Japanese

April 2017

Revision C

2410-902-01C

3.1 MB

- The full preview is over. If you want to read the whole 125 page document,please Sign in/Register -
  • +1 Like
  • Add to Favorites

Recommend

All reproduced articles on this site are for the purpose of conveying more information and clearly indicate the source. If media or individuals who do not want to be reproduced can contact us, which will be deleted.

Contact Us

Email: