Telaire® 7001 CO2 and Tempera ture Monitor
The Telaire® 7001 CO₂/Temperature monitor (shown in Figure 1 below) is an easy to use hand-held instrument, which provides stable and highly accurate readings due to Telairepatented dual beam NDIR technology. The monitor, equipped with a 0-4 V output, is ideal for long-term monitoring and recording.
Display Features:
•CO2 Readings (Upper Display) - Remain visible at all times
•Temperature and Ventilation Rates (Lower Display) - The Up/Down arrows allow you to toggle through the Temperature and Ventilation modes. When pressing the Up arrow, the display will go through the following sequence:Temp°C > Temp°F > Vent Ratel/p/s > Vent Ratecfm/p > Blank
•US Standard to Metric Conversion- The Temperature, Ventilation rates, and Elevation Readings can be viewed in US Standard or Metric Readings. The Temperature is converted from Fahrenheit (°F) to Celsius (°C); the Ventilation Rates are converted from Cubic Feet Per Minute Per Person (cfm/p) to Liters Per Second (l/p/s) and the Feet (ft) to Meters (m).
Press the up/down button until cfm/p is displayed. This value represents how much outside air is being introduced on a CFM per person basis. The reading is derived from calculating the outside air ventilation rate to a space based on the inside/outside CO₂ differential readings.
The current codes/standards generally require 15 to 20 CFM/Person to be delivered to most spaces to ensure acceptable air quality. Low values indicate low ventilation rates and potentially poor air quality. High levels indicate excessive ventilation and potential excessive energy usage. To obtain accurate measurements, readings should be taken 2 to 3 hours after occupancy has stabilized in a space or at a peak in daily CO₂ concentrations.
CO₂ and Temperature Monitor 、 easy to use hand-held instrument |
|
|
|
User's Guide |
|
|
|
Please see the document for details |
|
|
|
|
|
|
|
English Chinese Chinese and English Japanese |
|
September 2014 |
|
|
|
T62292-008 |
|
287 KB |
- +1 Like
- Add to Favorites
Recommend
All reproduced articles on this site are for the purpose of conveying more information and clearly indicate the source. If media or individuals who do not want to be reproduced can contact us, which will be deleted.