DSA1000A/DSA1000 Series Spectrum Analyzer Performance Verification Guide

2020-12-28

This manual guides users to correctly test the performance of RIGOLDSA1000A/DSA1000 series spectrum analyzer. For the operation method of the instrument, please refer to the corresponding User’s Guide
Test Preparations:
Before the test, make the following preparations.
1.Warm the spectrum analyzer up for at least 30 minutes.
2.Make sure that the instrument is within the calibration period (the calibration period is 1 year) and perform self-calibration on the spectrum analyzer (press System→Calibrate→Cal Now and the spectrum analyzer performs self-calibration immediately using the internal calibration source).
3.The test devices required are as shown in the table below.