DG4000 Series Function/Arbitrary Waveform Generator Calibration Guide
For optimum performance, all procedures should comply with the following recommendations.
1) Make sure the environment temperature is stable and between 18°Cand 28°C. Ideally, the calibration should be performed at 23°C±1°C.
2) Make sure the environment relative humidity is less than 80%.
3) Allow a 30-minute warm-up period before the calibration.
4) Keep the test cables as short as possible, consistent with the impedance requirements.
|
|