Agilent 81200 How to Use the 81200 Data Generator/Analyzer Platform together with Agilent VEE for Signal Integrity Analysis
●Introduction
■As designs get faster and more complex, signal integrity is becoming increasingly important within the design validation process.
■There are two main issues with signal integrity, these being waveform and jitter. Traditionally, this is an analog instrument domain, for products such as scopes.
■At a first glance, it sounds contradictory to use digital equipment to per-for analog type measurements, but we intend to prove otherwise. This product note explains how waveform measurements can be performed using Agilent Technologies ’81200data generator/analyzer together with Agilent VEE.
■Besides the basics, this product note will discuss, using an example program, how an output waveform(similar to that of a scope shot)is reconstructed on the 81200 with the help of Agilent VEE.This reconstruction is referred to as “soft scope ”later in the document.
■This method is not targeted to replace the traditional usage of an oscilloscope, as it still perfors this type of measurement much faster.Instead,VEE provides a more convenient way to achieve the same results, without the need to integrate a scope in to the measurement setup, thus saving time, money, and reducing setup complexity.
|
|