P5100 2500 V 250 MHz Oscilloscope Probe Instructions
●Features and Accessories
■The P5100 is a 100X, 2500 V (DC + peak AC), 1000 V CAT III oscilloscope probe that provides 250 MHz performance.
■The P5100 probe is designed for ground-referenced measurements only. The P5100 probe is only rated CAT I for voltages exceeding 1000 VAC and not intended for mains voltages above 1000 VAC. A CAT I circuit is a non-mains energy limited circuit within equipment. Refer to the Specification section for more details.
■The P5100 probe is compatible with all Class 1 (grounded) general-purpose oscilloscopes and with Tektronix ground-referenced oscilloscopes that automatically detect 100X probe attenuation and display the correct scale readout.
P5100 、 TDS724C 、 PG506A 、 SG503 、 011-0049-XX 、 103-0028-XX 、 013-0291-XX 、 012-0482-XX 、 TPS2000 、 THS700 、 206-0466-XX 、 200-3016-XX 、 200-3017-XX 、 200-3018-XX 、 206-0465-10 、 020-2137-XX 、 344-0461-XX 、 003-1433-XX 、 013-0290-XX 、 334-1275-XX 、 196-3424-XX 、 070-8151-XX 、 206-0060-XX |
|
|
|
User's Guide |
|
|
|
Please see the document for details |
|
|
|
|
|
|
|
English Chinese Chinese and English Japanese |
|
2018/06/21 |
|
|
|
070-8151-05 |
|
1.4 MB |