• I have my large slot (81640A/680A/642A/682A) tunable laser connected to an OSA. Why don‘t I see any change in the measured spectrum of the laser when coherence control is activated?

  • 创建于2020-05-08

1answer(s)

  • (0)
    Coherence control increases the linewidth of the laser from 100KHz to 500MHz(typical). At 1550nm this relates to a wavelength change of .004nm. The wavelength resolution of the OSA is .06nm resolution bandwidth minimum. It cannot resolve a .004nm change.