CAUTION: Do not ov er-tighten the connector . Ov er-tightening
can damage the center conductor .
5. F asten the 7mm connector of the test instrument onto the
RDL-A connector until snug (appro x. 3 turns).
6. Enter the electrical length to compensate f or fixture phase
dela y.
7. Make sure there is no component or shorting bar in the
fixture, and perf orm OPEN fixture compensation.
8. Place the supplied shorting bar onto the fixture. Center and
press the shorting bar into the contacts until completely
seated.
9. P erf orm SHORT fixture compensation. Remo ve the shorting
bar from the fixture.
10. Place the test component into the fixture.
11. Read the display ed value on the instrument.
References
The f ollowing application notes are av ailable on the Coilcraft
website at
www.coilcraft.com/appnotes.cfm
Test Fixture Compatibility Chart
Calibration, Compensation and Correlation