CA UTION: Do not ov er-tighten the connector . Ov er-tightening
can damage the center conductor .
5. F asten the 7 mm connector of the test instrument onto the
AXL-A connector until snug (appro ximately 3 turns).
6. Enter the electrical length to compensate f or fixture phase
dela y.
7. Make sure there is no component or shorting bar in the
fixture, and perf orm OPEN fixture compensation.
8. Place the supplied shorting bar into the fixture. Press the
shorting bar into the contacts until completely seated.
9. P erf orm SHOR T fixture compensation. Remov e the shorting
bar from the fixture.
10. Place the test component into the fixture.
11. Read the display ed value on the instrument.
References
The f ollowing application notes are av ailable on the Coilcraft
website at
www.coilcraft.com/appnotes.cfm
Test Fixture Compatibility Chart
Calibration, Compensation and Correlation
Testing Inductors at Application Frequencies
®