This standalone data logger measures and stores more than 32,000 temperature readings over a -40 to +125°C (-40 to +257°F)
range with a resoluon of 0.1°C (0.2°F).
The user can easily set up the logger and view downloaded data by plugging the data logger into a PCs USB port and using the free
EasyLog 21CFR soware. Data can then be graphed, printed and exported to other applicaons for detailed analysis. The encrypted
data has full audit tracking to comply with the requirements of 21CFR Part 11.
The high contrast LCD can show a variety of temperature informaon. At the touch of a buon, the user can cycle between the
current, maximum and minimum stored values for temperature.
The data logger is supplied with a lithium metal baery which typically gives two years’ logging life.
▪ -40 to +125°C (-40 to +257°F) measurement range
▪ Stores over 32,000 readings
▪ EasyLog 21CFR soware available as a free download
▪ Logging rates between 1 second and 12 hours
▪ High contrast LCD, with four digit temperature display funcon
▪ Immediate, delayed and push-to-start logging
▪ User-programmable alarm thresholds
▪ Use as part of a 21CFR Part 11 compliant system
Measurement range -40 to +125°C (-40 to +257°F)
Accuracy (logger error) ±0.1°C (±0.2°F)
Accuracy (probe) See ‘Temperature Probe Accuracy’ on page 4
Resoluon (display) 0.5°C (1°F)
Resoluon (data) 0.1°C (0.2°F)
Logging rate User selectable between 1 second & 12 hours
Operang temperature range -35 to +80°C (-31 to +176°F) (data logger only)
Baery Life 2 years (at 25°C and 1 minute logging rate,
LCD on)
Readings 32,510
Dimensions 135 x 24 x 21mm (5.31 x 0.94 x 0.82”)
SPECIFICATIONS ACCESSORIES
INCLUDED IN THE BOX
Replacement baery
Baery
Mounng Bracket
Replacement standard
thermistor probe
Standard thermistor
probe
High accuracy
thermistor probe
Probe extension
(5.0m)
Probe extension
(10.0m)
BAT 3V6 1/2AA
BAT 3V6 1/2AA
EL-LCD WALL BRACKET
EL-P-TP
EL-P-TP
EL-P-TP+
EL-PROBE-EXTENDER-
5.0M-TP
EL-PROBE-EXTENDER-
10.0M-TP
CALIBRATION CERTIFICATES NOW AVAILABLE
Lascar now oers a Traceable Calibraon Cercate Service on Temperature Data Loggers. Using reference equipment
which has been calibrated by a UKAS/NIST accredited laboratory and using apparatus traceable to naonal or
internaonal standards. For more informaon, please see www.lascarelectronics.com.
EL-21CFR-TP-LCD
EasyLog 21CFR Compable
Temperature Probe Data Logger with LCD Screen
www.lascarelectronics.com
Issue 3 10/2016 Page 2 of 5
EL-21CFR-TP-LCD
EasyLog 21CFR Compable
Temperature Probe Data Logger with LCD Screen
Easy to install and use, Lascar’s EasyLog 21CFR soware is compable with all latest versions of Windows (7, 8 & 10 - both 32-bit &
64-bit) and is available as a free download from www.EasyLog21CFR.com. All data collected from the logger and associated audit
trails are stored in an encrypted format which cannot be edited.
For more informaon, and to download the latest version of the soware free of charge, visit www.EasyLog21CFR.com
STORE DATA IN COMPLIANCE WITH THE REGULATIONS OF 21CFR PART 11
CONTROL YOUR LOGGER
Users can congure their loggers with the following parameters:
▪ Logger name
▪ Temperature measurement parameter (°C or °F)
▪ Logging rate (user selectable between 1 second and 12 hours)
▪ High and low alarms for temperature
▪ Immediate and delayed logging start
▪ Probe type
Once users have recorded data, the built-in graphing soware allows them
to graph and annotate their data, or export it to Excel, PDF or jpeg formats.
CONTROL YOUR DATA
EasyLog 21CFR soware ensures digital security and compliance:
▪ Assign individual users with specic permissions
▪ Full soware & session data audit trails
▪ Receive email alerts for failed log in aempts
▪ Digital signatures added to all reports
▪ Add comments to specic readings