1
QM5 Series
QM5FM_Datasheet_260323 - v14.0
QM5 Series
Medical
700W / 1200W
Modular power supply
Features Benets
• BF ready medical isolation (MOPP) Eases design into systems (including BF)
• Low speed, low audible noise fan Enhanced patient / user experience
• Up to 12 outputs Eliminates need for additional supplies
• PMBus™ communication option Remote monitoring and control
• 7 year warranty
Low cost of ownership
Industrial
Test
Broadcast
Comms
Renewable
Input QM5 QM5H
Output power 700W 800W 700W 1200W
Input voltage 90-264Vac 180-264Vac 90-264Vac 180-264Vac
Frequency 47 - 63 Hz (440Hz with reduced PFC)
Input fuses 16A / 250Vac HBC Fast acting (not user accessible) in both Live and Neutral lines (single fusing optional)
Inrush current <40A at 25°C and 264Vac (cold start)
Leakage current See ‘How To Create A Product Description’ for details
Touch current <100µA (with 4 or fewer modules). For other congurations, contact sales for details.
Power factor > 0.95 (at 230Vac, 100% load)
How To Create A Product Description
The extensive range of output modules and options make it possible to achieve almost any combination of Volts and Amps. You can create
your own QM conguration online at https://cong.emea.tdk-lambda.com/. This method checks your conguration and oers the optimum
solution. Alternatively, you can do this manually by using the guide below.
1. Calculate total output power to select the appropriate converter, then select required Cooling, Connection, Leakage Current and Controls/
Signals from the following table:
2. Select Output Modules using the output voltages tables and the module specications.
3. Contact TDK-Lambda to validate conguration and issue a part number.
Converter
5 slots wide
QM5
700W low line, 800W high line
QM5H
700W low line, 1200W high line
QM5 F S D L T5H
Cooling
F Variable speed Forward air - standard
R Reverse air (contact sales for details)
ustomer air - no fan (contact sales for details)
Input
Connection
S Screw
I I
EC320
(max at 264Vac, 63Hz)
Must also add the leakage current from
L 212µA
R 77µA
Input fuse
E
Single AC fuse in Live line
(contact sales for details)
Standby
/ Signals
none
E5L 5V / 250mA, Enable
E5H 5V / 2A, 5V / 250mA, Enable
E12H 12V / 1A, 5V / 250mA, Enable
T5L 5V / 250mA, Inhibit
T5H 5V / 2A, 5V / 250mA, Inhibit
T12H 12V / 1A, 5V / 250mA, Inhibit
P5H
5V / 2A,
see PMBus™ app note
see speci cation page for details
Isolation
Input to output / signals Reinforced
2 x MOPPs (3rd edition 60601)
4kVac, 5.7kVdc type tested to 4kVac (equivalent to 5.7kVdc), production tested to 4.3kVdc.
Input to earth Basic 1 x MOPP, 1.5kVac
Output / signals to earth Basic 1 x MOPP, 1.5kVac
Output / signals to output / signals Basic 200Vdc
https://emea.lambda.tdk.com/qm
https://product.tdk.com/en/power/qm