Product Brochure | Version 05.00
R&S®ENV216
Two-Line V-Network
For disturbance voltage
measurements on
single-phase EUTs
The R&S®ENV216 two-line V-network meets the
requirements of CISPR 16-1-2 and EN 55016-1-2
for V-networks with a simulated impedance of
(50µH +5Ω) || 50Ω in the frequency range from
9kHz to 30MHz as well as the ANSI C63.4 in the
frequency range from 150 kHz to 30 MHz. A female
connector with protective earth contacts is pro-
vided for connecting the EUT. Various models with
country-specific connector systems are available.
Owing to its compact design and low weight, the
R&S®ENV216 is ideal for frequently varying applications
and especially for on-site disturbance voltage measure-
ments on single-phase EUTs. Note that high leakage
currents are produced because of the standard design
of these V-networks. The networks should therefore be
connected to a low-impedance protective earth system. In
uncertain cases, an isolating transformer should be used.
In the case of AC supply operation in a range from
90 V to 240 V and 50 Hz to 60 Hz, the operating voltage
for the built-in logic circuit is obtained from the AC supply
direct. This covers all standard applications. If the EUT is
operated with low AC voltages or DC voltages up to 50 V,
the logic circuit is powered via an external plug-in power
supply that is supplied as standard.
Key facts
Frequency range 9 kHz to 30 MHz
Power-handling capacity up to 16 A, constant current
(country-specifi c)
Simulated impedance (50 µH + 5 Ω) || 50 Ω in line with
CISPR 16-1-2 Amd. 2:2006
V-network in line with CISPR, EN, VDE, ANSI,
FCCPart15 and MIL-STD-461D, E and F
Calibrated in line with CISPR 16-1-2
2
R&S®ENV216
Two-Line V-Network
At a glance