1
Instrument Software Update Instructions
VXG and VXG-m Vector Signal Generators
M9383B, M9384B
Purpose
The purpose of this document is to outline the instrument software update procedure from the revision
that is currently installed, to the latest revision available. The advantages of updating the instrument
software with the current revision include:
1. Keeps the instrument current and up to date
2. Updates the generator applications where appropriate
3. Updates the PC board software where appropriate
An update to the instrument software version does not require a new license key for the generation
application(s), so long as the application(s) were licensed prior to the update.
This procedure will not update any other software.
Note: Connecting a mouse and keyboard is recommended for the M9384B VXG. For the M9383B
VXG-m a mouse, keyboard, and monitor are required or use remote access via Microsoft Windows
Remote Desktop.
To determine the revision of instrument software currently installed, tap the gear icon on the top
bar of the graphical user interface, then tap System. Look for the Firmware Versionas shown
below. If the installed revision is the same as the latest revision on the web, there is no need to
update the software.
2
Instrument Software Installation
Requirements
To use this update procedure, the following requirements must be met:
Instrument must be functioning properly before the update is attempted
The ability to login to the instrument as the administrator
The ability to transfer a large file (~1.5 GByte) to the instrument being updated
Obtaining the Latest Software Update
1. Go to http://www.keysight.com/find/vxg
2. Select the appropriate link in the Software Update section of the table
3. Download the executable file (Keysight.PathWave.SG.Setup_A.XX.XX.exe) by clicking on
the Download button. Save the file to a location of your choice; recommendation is to use a
folder on the D: partition.
If you do not have the ability to download files directly to the D: partition of the instrument being
updated, store the Keysight.PathWave.SG.Setup_A.XX.XX.exe file on a large capacity USB storage
device for transfer to the instrument. (The file size is ~1.5 GByte.)
Note: This will update the instrument software currently running on the instrument. It will not update
the version that exists on the recovery partition. Because of this, if the recovery process is performed,
the instrument will revert to the version that was originally shipped from the factory. It is
recommended you place a copy of this software installer on the D: partition, so you can easily perform
an update in the event you need to do a recovery in the future.
Update Procedure
The default user account is Instrument, which does not have the required permissions to install the
instrument software updates. For the process outlined below, ensure the Instrument user is Signed
Out, do not use Switch User, then login as the administrator.
The instrument software upgrade process has an install wizard which removes the old software
version and installs new software version. The time to perform the upgrade will may vary between 10
and 45 minutes depending if FPGA upgrading is performed. The FPGA updates are handled
automatically during the instrument software update process.
1. If the instrument is not already running, turn the instrument on and allow it to completely boot
up
2. Close the instrument application by pressing the icon on the top bar of the graphical user
interface, then Exit