DQD 507 Rev. 2019-04-30 © 2018 CSA Group. All rights reserved. Page 1
Cer t i ficate of Compliance
Certificate:
80008669
Master Contract:
255632
Project:
80008669
Date Issued:
2019-12-03
Issued To:
Analog Devices Incorporated
804 Woburn St.
Wilmington, Massachusetts, 01887-3462
United States
Attention: Keith Bennett
The products listed below are eligible to bear the CSA Mark shown with
adjacent indicators 'C', 'US' and
Issued by:
Martin Buchanan
Martin Buchanan, P. Eng.
PRODUCTS
Component Acceptance of Isolator Module Devices:
Device
(BG A L GA 44 package)
Ratings
Clauses of Standard/Notice
Internal
External
kV ˚C
Creepage
(mm)
Dist
Thru
(mm)
Creep/Clear
(mm)
LTM2884wx
3.0
105
CSA
60950-1-07+A1+A2 2.10.3.3, 2.10.4.2,
2.10 .4.3, 2.10.5 .4a, 2.10.11 , 4. 5.2, 5.2
62368-1-14 5.4.3, 5 . 4.2, 5.4.4.4, 5.4. 7,
5.4. 1.5.3, 4.5.8, 5.4.9.1, 4.1.2
60601-1:14 8. 5.5.1, 8.8.2 , 8. 8.3x1.6, 8.9.3.2,
8.9. 3.4, 8.9.1.1 5, 8.9.1. 7
-
-
9.3
Certificate: 80008669
Project: 80008669
Master Contract: 255632
Date Issued: 2019-12-03
DQD 507 Rev. 2019-04-30 © 2018 CSA Group. All rights reserved. Page 2
Device
(BG A L GA 44 package)
Ratings
Clauses of Standard/Notice
Internal
External
kV ˚C
Creepage
(mm)
Dist
Thru
(mm)
Creep/Clear
(mm)
IEC
60950-1 2nd Ed.,+A1+A2 2.10.3.3,
2.10 .4.2, 2.10.4 .3, 2.10.5.4 a , 2.10. 11 , 4.5.2, 5.2
62368-1-14 5.4.3, 5 . 4.2, 5.4.4.4, 5.4. 7 ,
5.4. 1.5.3, 4.5.8, 5.4.9.1, 4.1.2
60601-1 Ed.3+A1 8.5.5.1, 8.8.2,
8.8. 3x1.6, 8. 9 . 3 .2, 8.9.3.4, 8.9.1.15, 8. 9.1.7
UL
UL 1577-2015 Standard for Optical
Isolators - Single protection
where w is a letter for temperature range up to 105C, x is V for LGA package or Y for BGA package.
Notes:
1. These devices meet basic insulation requirements for 930Vrms for CSA 60950-1-07+A1+A2 and IEC 60950-1
2
nd
Ed.+A1+A2. (pollution degree 2, material group III)
2. These devices meet reinforced insulation requirements for 415Vrms for CSA 60950-1-07+A1+A2 and IEC
60950-1 2
nd
Ed.+A1+A2. (pollution degree 2, material group III)
3. These devices meet basic insulation requirements for 600Vrms for CSA 62368-1-14 and IEC 62368-1:2014.
(pollution degree 2, material group III)
4. These devices meet reinforced insulation requirements for 300Vrms for CSA 62368-1-14 and IEC 62368-
1:2014. (pollution degree 2, material group III)
5. For CSA 60601-1:14 and IEC60601-1 Ed.3+A1 for 2MOPP for 150Vrms, the devices meet clauses 8.5.5.1,
8.8.2, 8.8.3x1.6, 8.9.3.2, 8.9.3.4, 8.9.1.15, 8.9.1.7. The risk management process is not applicable to these clauses.
6. Case material: CTI=600V, erosion depth 0.028mm, meets material group I.
7. The creepage and clearance has been evaluated for altitudes 2000m, in pollution degree 2, material group III
and overvoltage category II except where specified otherwise.
These devices are Component Accepted as components for use in other Certified equipment where the suitability
of the combination shall be determined by investigation in the final application.
APPLICABLE REQUIREMENTS
CAN/CSA-C22.2 No 60950-1-07+A1+A2 - Information Technology Equipm ent - Safety - Part 1: General
Requirements (Bi-national Standard, with UL 60950-1)
CAN/CSA-C22.2 No. 62368-1-14 - Audio/video, information and communication technology
equipment– Part 1: Safety requirements
IEC 60950-1 2
nd
Ed.+A1+A2 - Information Technology Equipm ent - Safety - Part 1: General Requirements