Application Note
PulseView
Calibrations
Introduction
The PulseView mode in the MS464XB VNAs allows point-in-pulse, pulse profiling and pulse-to-pulse measurements and
often these are performed in some calibrated sense beyond simple normalization. While these are still S-parameter
measurements and all of the usual algorithms (SOLT, SOLR, SSLT, TRL, LRL, LRM…) and types (full 2-port, 1 path-2 port,
full 1 port, frequency response…) still apply, there can be some setup and configuration differences to be aware of.
The purpose of this note is to discuss these setup and configuration questions in the context of some common pulse
measurement scenarios.
MS4640B Series Vector Network Analyzer
2
Basic configurations
The calibration implications are heavily dependent on the basic measurements configuration. Is pulsed RF stimulus
being used? If not (e.g., the pulsing is only of the DUT bias or some control signal), then by definition, the calibration
will be performed in a non-pulsed state. When performing pulse measurement calibrations, one must be in the
PulseView application as calibrations cannot be shared between regular Transmission/Reflection and PulseView
since different IF systems are used. While using the PulseView application, the desired mode (Point-In-Pulse (PIP),
Pulse Profiling (PP) or Pulse-to-Pulse (P2P)) must be selected. The latter is an important point since the x-axes are
not the same (frequency vs. time vs. time/number of pulses). Since this calibration is non-pulsed, the Point-In-Pulse
situation is conceptually simple…a frequency sweep (or power sweep) is performed and appropriate error coefficients
generated. Changing the measurement width or position (or IFBW) after calibration will only affect relative levels of
trace noise and dynamic range of the results. The frequency range and point count are constrained as in regular
transmission/reflection mode; although interpolation is allowed. For Pulse Profile and Pulse-to-Pulse, the calibration
is essentially a CW process since the same frequency and power are used for the measurements across an axis of
time…the same error coefficients, within limits of trace noise, will be generated at each point and applied to the
measurement result. Again, measurement widths and positions can be changed after the calibration with only relative
changes in trace noise or dynamic range. The number of points, however, cannot be changed as this would alter
the indexing of the error coefficients.
If stimulus pulsing is needed (using Anritsu pulsed test sets or with external modulators), the situation can be slightly
more complex. The central issue is if and to what extent one wants to calibrate out dynamics associated with the
stimulus modulation process. One aspect to consider is how the reference signals (a1, a2…) are handled. The Anritsu
test sets (SM68XX) have a post-modulator reference coupler that allows pulsed references to be used (in lieu of the
unpulsed reference coupled internally to the VNA). Alternatively, a non-pulsed reference can be used by not connecting
these paths (and both methods can be emulated using external modulators). The methods for connecting the test
set in these two alternative methods are shown in Figure 1. The resulting S-parameters are different in these two
cases as suggested in Figure 2 for a PP example. This is an uncalibrated measurement just to show the behaviors
of the ratio and, since the paths of the reference are different in the two cases, the absolute S21 value changes
slightly. The central point is again the transition behavior.
Figure 1. Setups using the SM68XX pulsed test set with (picture on left) and without (picture on right) pulsed references are shown here
(cable changes circled). Similar cable changes occur on the rear panel for frequencies less than 2.5 GHz.