4. Always place the X-Cite unit close enough to the microscope so that there is some slack in
the light guide and no sharp bends.
5. Never leave an endcap on the output end of the light guide when the other end is connected
to the X-Cite unit; if the unit is turned on in this condition, the cap will overheat, melt and/or
permanently discolor the quartz end of the light guide.
6. Do not expose the light guide to extreme temperatures (above 35°C, below -5°C) for
extended periods of time during use, transport or storage; this may cause degradation of the
seals and allow air bubbles to form in the liquid.
What does a bubble look like?
Depending on the size and location in the light guide, a bubble may or may not be obvious. To
check for bubbles:
1. Disconnect the LLG from the X-Cite and microscope adapter.
2. Hold one end towards a bright window or overhead room light - DO NOT use an X-Cite or
any other focused light source for this test!
3. Look at the quartz at the other end of the LLG
a. Bubble-free: quartz end will appear as a bright, solid circle; you may also be able to
see a thin circular outline at the quartz/liquid interface.
b. Bubbles at/near the quartz end: appear as dark spots, as small as 0.5mm in
diameter or even as larger more defined spheres.
c. Bubbles in the middle of the light guide: may not be well-defined spots, but will
appear as dark shadows
d. In extreme cases, where the bubble is blocking the entire diameter of the light guide,
no light will come through, even when pointing the distal end at a light source.
100% 11.3% 5.2%