1
Transmitter
BiT – Barksdale Intelligent Transmitter Series H455E, 455E
Features
Intrinsic safety certification with ATEX, IECEx and CSA approvals
Lightweight, compact and all-welded rugged stainless steel construction
High performance sensors
Optional HART® communication protocol with diagnostics & field
calibration, reducing maintenance time and cost; 10:1 turndown ratio
for re-ranging, maintaining high accuracy and allowing one part number for
multiple pressure ranges
High accuracy: ±0.1% and 0.25% FSO, (L,H,R)
Additional CSA certification for ordinary locations
Factory sealed IP66 & IP67 and NEMA 4X enclosure
CE, NACE compliant and superior EMC/EMI protection
RoHS / REACH Compliant
General Specifications*
* See product configurator for additional options.
** Consult factory for 316SS option
Accuracy
including Linearity,
Hysteresis and
Repeatability:
at 75°F, Typical
±0.1% of Calibrated FSO (Option [–A1])
±0.25% of Calibrated FSO (Standard)
±0.5% of Calibrated FSO (Vacuum only
[-23])
Temperature Shift:
Zero & Span: 0.0125% Per °F over the
compensated temperature range
Long Term
Stability:
±0.2% FSO /year of calibration curve
Typical Life Cycle:
100 million cycles
Proof Pressure:
2X range for up to 7500 psi models
1.5X range for 7500 - 30000 psi models
Input:
Excitation voltage: 9 to 30 VDC
Supply Current:
20mA max.
Output:
H455E:
455E:
Output: 4–20 mA with HART® Protocol
Full Scale Output: 20 mA ±1%
Zero output: 4 mA ±1%
Output: 4–20 mA
Full Scale Output: 20 mA ±1%
Zero output: 4 mA ± 1%
Dynamic
Response Time:
<70 milliseconds
Enclosure:
All welded 300 series stainless steel.
IP66 & IP67 and NEMA 4X rated. 316
Stainless Steel (optional)
Wetted Parts:
Sensor:
Fitting:
17-4 PH Stainless steel (more than 10k psi)
316L Stainless steel (up to 10k psi)
Inconel (optional: up to 10k psi)
17-4 PH Stainless steel for -UL option (10k psi)
17-4 PH Stainless steel (all ranges)
316L Stainless steel and 718 Inconel (optional: up to
10k psi)
Media
General:
Corrosive and Acidic:
(up to 10,000 psi)
Gas, vapor, liquid and viscous fluids [-Z17]
718 Inconel wetted material, NACE [-NC], and 316L
stainless steel, NACE [-SS]
Pressure Connection:
13 available options. Refer to product configurator
for available connections and ranges.
Electrical Connection:
10 available electrical connections such as M12,
Bendix, DIN, conduit, gland and etc. Option for free
leads from conduit or gland electrical connection
includes two signal leads plus ground lead
consisting of 18AWG halogen free type, 80” (2m)
long or optional jacketed cable exiting 1/2 NPT or
M20 conduit or gland connections.
Temperature Ranges:
Operating (Ambient):
Compensated:
Media:
Storage:
-40 to +176 °F (-40 to +80 °C)
0 to +165 °F (-18 to +74 °C)
-40 to +176 °F (-40 to +80 °C)
-40 to +185 °F (-40 to +85 °C)
Vibration:
10 g’s, 10-500 Hz, MIL-STD 202, Method 204, Cond. A
Shock:
50 g’s, 11 mS, MIL-STD 202 Method 213, Cond. G.
* See product configurator for additional options.
Applications
Drilling rig control & monitoring instrumentation
Drilling and frac offshore equipment skids
Production control systems
Chemical injection systems
Oil & gas pipelines
Petrochemical plants
Refineries
Gas and oil fired power plants
Hydraulic power units
Gas transfers for fuel systems
O&G process pumps
Gas compressors
Generators and turbines
Subject to technical changes.
2
BiT – Barksdale Intelligent Transmitter Series H455E, 455E
Approvals:
CSA
CSA
ATEX/IECEx:
cCSAus Ordinary Locations
cCSAus Intrinsic Safety for Hazardous
Locations
Class I, Div. 1, Groups A, B, C & D
Class II, Div. 1, Groups E, F & G
Ex ia IIC T4 Ga
Ex ia IIIC T135°C Da
Class 1 Zone 0, AEx ia IIC T4 Ga
Zone 20 AEx ia IIIC T135°C Da
Factory sealed IP66 & IP67 and NEMA 4X
enclosure
CE 0081
II 1 G D
Ex ia IIC T4 Ga
Ex ia IIIC T135°C Da, IP66 & IP67
-40 °C Tamb +80 °C
CML 19 ATEX 2240X
IECEx CML 19.0066X
HART® Protocol:
Turndown Ratio:
Analog Output
Calibration:
Sensor Trim:
Diagnostics:
HART® 7.6
10:1 (Full scale output rangeability)
Adjustable zero and span of 4-20mA.
Full scale range must be (calibrated
span / 10)
Sensor zero and full scale trim
Transmitter failure indicated by off-scale
analog signals to alarm the user
See HART Installation and Operation
Manual 272438 for detailed information
Compliances:
CE,
NACE (316L SS or 718 Inconel wetted
material)
Electromagnetic
Compatibility (EMC)
IEC/EN 55011:
IEC/EN 61000-4-2:
IEC/EN 61000-4-3:
IEC/EN 61000-4-4:
IEC/EN 61000-4-5:
IEC/EN 61000-4-6:
Emission & radiated emission for class A
limits
Electrostatic discharge (ESD) test - contact
discharge +/-4 kV, Air discharge +/-8 kV
Radiated RF, EM field immunity
80mhz-1ghz, 3v/M
EFT (Burst) Test, +/-2 kV
Surge Test, +/-1kV between line and earth
ground
RF Immunity, 150 kHz – 80 MHz, 3V
Weight:
16 Ounces (453 grams)
Warranty:
1 year warranty
Subject to technical changes.
* See product configurator for additional options.
General Specifications* cont.
Wiring Code
TABLE 1. FREE LEAD WIRES AND CONNECTOR PIN CONNECTIONS
MODEL NO. RED/A/1 BLACK/B/2 GREEN/D/4
H455E + EXCITATION - EXCITATION EARTH GROUND
455E + EXCITATION - EXCITATION EARTH GROUND
Installation: Intrinsic Safety Installation for Hazardous Locations:
Install per Barksdale Control Drawing 272471
I.S. Entity
Parameters
Supply/Signal Barrier PressureTransmitter
Vmax (Ui) Imax Ci Li Pi max
30 VDC 100 mA 2.2 nF 30 μH 750 mW
TABLE 2. JACKETED CABLE WIRE CONNECTIONS (-J Option)
MODEL NO. WHITE BROWN GREEN
H455E + EXCITATION - EXCITATION EARTH GROUND
455E + EXCITATION - EXCITATION EARTH GROUND
CONFIGURATION WITH HART®
Point to Point
+
_
+
_
METER
BLK(-EXC)
RED(-EXC)
POWER SUPPLY
9 - 30 VDC
CASE
GROUND
455
TRANSDUCER
CURRENT OUTPUT
WIRING
A
V
WHT (+ OUT)
VOLTAGE OUTPUT
452
TRANSDUCER
CASE
GROUND
POWER SUPPLY
11/14-30 VDC
RED(+ EXC)
BLK
(COMMON)
_
+
_
+
+
_
+
_
METER
BLK (-EXC)
RED(+ EXC)
POWER
SUPPLY
425
TRANSDUCER
CURRENT OUTPUT
A
LOOP RESISTANCE
REQUIRED (OHMS)
EXCITATION VOLTAGE (volts DC)
OPERATING
BAND
MAXIMUM
RESISTANCE
363025201512
0
300
600
900
1200
R
250
Black
Red
-
+
HART Master /
Handheld Communicator
H455E
Point to Point :
POWER SUPPLY
Red
Black
Red
Black
H455E
#15
H455E
#2
Black
Red
H455E
#1
Multi Drop :
HART
MASTER
R
250
Min 250R
@ 20mA
1500
9
+ +
+
+
-
-
-
-
CASE
GROUND
Green
GRN GRN
Green
Input Card/ Module
Field Operation:
H455
+
-
Red
Black
Multi Drop
+
_
+
_
METER
BLK(-EXC)
RED(-EXC)
POWER SUPPLY
9 - 30 VDC
CASE
GROUND
455
TRANSDUCER
CURRENT OUTPUT
WIRING
A
V
WHT (+ OUT)
VOLTAGE OUTPUT
452
TRANSDUCER
CASE
GROUND
POWER SUPPLY
11/14-30 VDC
RED(+ EXC)
BLK
(COMMON)
_
+
_
+
+
_
+
_
METER
BLK (-EXC)
RED(+ EXC)
POWER
SUPPLY
425
TRANSDUCER
CURRENT OUTPUT
A
LOOP RESISTANCE
REQUIRED (OHMS)
EXCITATION VOLTAGE (volts DC)
OPERATING
BAND
MAXIMUM
RESISTANCE
363025201512
0
300
600
900
1200
R
250
Black
Red
-
+
HART Master /
Handheld Communicator
H455E
Point to Point :
POWER SUPPLY
Red
Black
Red
Black
H455E
#15
H455E
#2
Black
Red
H455E
#1
Multi Drop :
HART
MASTER
R
250
Min 250R
@ 20mA
1500
9
+ +
+
+
-
-
-
-
CASE
GROUND
Green
GRN GRN
Green
Input Card/ Module
Field Operation:
H455
+
-
Red
Black
TABLE 3. DEUTSCH CONNECTOR PIN CONNECTIONS (-D3 & -D4 Option)
MODEL NO. CONNECTOR PIN A/1 PIN B/2 PIN C/3 PIN D/4
H455E & 455E D3 + EXCITATION - EXCITATION EARTH GROUND N/A
H455E & 455E D4 - EXCITATION + EXCITATION EARTH GROUND N/A
4 53/64"
122.7
5 21/64"
135.4
.40
10.2
1/4"
6.4
FACTORY SEALED HALOGEN FREE
18 AWG LEADS 80 INCHES (2 METERS)
REFER TABLE 1 & 2.
OPTIONAL:
JACKETED CABLE
(LEADS/CABLE: BLANK, -T6, -H3)
ELECTRICAL CONNECTIONS:
STD(BLANK):
1/2"-14 NPT MALE
CONDUIT THREADS
OPTIONAL:
T2:
BENDIX 6P, SS HERMETIC
T3
: BENDIX, 6P, ALUM. FLANGE
T4:
M12 (4 Pin) CONNECTOR
T5:
STD DIN (43650)
T6:
M20X1.5 THD
T7:
M8 (4 Pin) CONNECTOR
T8:
4 PIN BENDIX, SS HERMETIC
H3:
WATERTIGHT GLAND WITH
18 AWG FREE LEADS
H4:
SUBMINATURE DIN (43650)
.03 OR Ø3/16
PRESSURE PORT OPENING
A/F
1" or 1 1/8"
[25.4] or [28.6]
1/4 NPT FEMALE THD
(-P6 SHOWN)
REFER CONFIGURATION
TABLE ON SHEET 2 FOR OPTIONAL
PROCESS CONNECTIONS.
1 3/8"
35
5"
127.7
.59
15
14
-T2 TERMINATION:
BENDIX CONNECTOR
MS3113 TYPE ,
6 PIN, SS HERMETIC
STD, -T6 MODEL
A
B
C
D
E
F
'A' TO 'F' ARE PINS
FOR CONNECTION
REFER TABLE 1.
-T2 MODEL
Copyright © 2017-19, Barksdale Inc., All Rights Reserved
18
STANDARD MODIFICATIONS: ADD IDENTIFIER TO P/N
(EXAMPLE: H455-01-W180 = STANDARD 0-15 PSI HART MODEL WITH 180 INCH LONG WIRES).
17. TYPE 4X ENCLOSURE ENVIROMENTAL RATING IP66 & IP67.
16
ALL PRESSURE RANGES HAVE BUILT-IN PRESSURE SURGE PROTECTOR (SNUBBER).
ADD "-Z17" SUFFIX FOR HIGH PRESSURE MODELS WITH NO SNUBBER.
15
PART NUMBERS: ARE RATED IN PSIS. FOR RATING IN PSIA INSERT "-A" FOLLOWING
THE PRESSURE RATING. (EXAMPLE: H455-04-A IS RATED IN PSIA, H455-04 IS RATED IN PSIS)
14
BENDIX CONNECTOR (-T2) OPTION.
PIN D, E, F ARE RESERVED FOR SPECIAL MODELS ONLY.
13. MATERIAL OF CONSTRUCTION:
HOUSING, ELECTRICAL CONNECTION:
304 STAINLESS STEEL, OPTIONAL: 316 STAINLESS STEEL
.
12. WETTED MATERIAL:
SENSOR:
STANDARD: THIN-FILM SENSOR, 17-4 PH STAINLESS STEEL
(15k-30k PSI AND 10k PSI WITH -UL OPTION) MODELS
STANDARD: DIFFUSED SILICON SENSOR, 316L STAINLESS STEEL (UP TO 10k PSI) MODELS
OPTIONAL: THIN-FILM SENSOR, 718 INCONEL (UP TO 10k PSI) MODELS
FITTING (PROCESS CONNECTION):
STANDARD: 17-4 PH STAINLESS STEEL (15k-30k PSI)
OPTIONAL: 316L STAINLESS STEEL (UP TO 10k PSI)
OPTIONAL: 718 INCONEL (UP TO 10k PSI)
11. THERMAL, ZERO AND SPAN EFFECT:
0.0125% PER
F OVER THE COMPENSATED TEMPERATURE RANGE.
10. SPAN ACCURACY:
1% FSO AT 75
F (24
C).
9. SPAN ADJUSTMENT (TURN DOWN): 10:1 (WITH HART OPTION)
8. ZERO BALANCE:
1% FSO AT 75
F (24
C).
7. RESPONSE TIME: 70ms
6. COMPENSATED TEMPERATURE RANGE: 0 to +165
F (-18 to +74
C)
STORAGE TEMPERATURE RANGE: -40 to +185°F (-40 to +85°C)
MEDIA TEMPERATURE RANGE: -40 to +176°F (-40 to +80°C)
OPERATING (AMBIENT) TEMPERATURE: -40 to +176
F (-40 to +80°C)
5. ACCURACY (LINEARITY, HYSTERESIS, REPEATABILITY):
0.25% FSO AT 75
F.
±0.10% FSO AT 75
F (OPTIONAL).
±.5% FOR VACUUM ONLY [-23] TYP.
4
DIGITAL AMPLIFIER (CURRENT/VOLTAGE OUTPUT):
H455, H455E: 4 - 20 mA (WITH HART OPTION)
455, 455E: 4 - 20 mA
452: 1 - 5 VDC
3
EXCITATION VOLTAGE:
H455, H455E: 9 TO 30 VDC (WITH HART OPTION)
455E: 9 TO 30 VDC
455: 9 TO 30 VDC (CLASS II POWER SUPPLY)
452: 12 TO 30 VDC (CLASS II POWER SUPPLY)
2
PROOF PRESSURE: 2X RANGE FOR UP TO 7500 PSI MODELS;
1.5X RANGE FOR 7500 - 30000 PSI MODELS.
1
PRESSURE RANGE: REFER PRODUCT CONFIGURATOR ON PAGE 2.
NOTES: UNLESS OTHERWISE SPECIFIED
29
22
REV.
DATE
SCALE
WEIGHT
NEXT ASSEMBLY
DWG.
REV.
#
C
DIST.
GmbH
REFERENCE DPM
PCO
1
2
3
4
A
B
C
D
4
3
2
1
D
C
A
B
1:1
SHEET 1
OF 3
-
G
Los Angeles, CA 90058-0843
Ph: 323-589-6181 Fax: 323-583-6252
DRAWN DATE
PAVAN
01-23-2018
CHECKED DATE
APPROVED DATE
UNLESS OTHERWISE
SPECIFIED
DIMENSIONAL TOLERANCES
FRACTIONAL ±1/16
DECIMAL ±.010
ANGULAR ± 5°
DIMENSIONS IN BRACKETS
ARE IN MILLIMETERS
www.barksdale.com
450
BARKSDALE INTELLIGENT TRANSMITTER
WITH OPTIONAL HART COMM PROTOCOL
A.4
03-19-2018
02-23-2018
X04
X01
01-23-2018
XX-XX-2017
YONG K
982217, 982218,
982219, 982220,
982434, 982435
DPM-4202.034
-----------------------
X02
01-24-2018
X03
02-05-2018
02-28-2018
A.1
03-05-2018
A.2
A.3
03-09-2018
03-27-2018
A.5
04-02-2018
A.6
04-16-2018
A.7
A.8
04-23-2018
A.9
05-08-2018
05-11-2018
A.10
A.11
07-17-2018
A
07-24-2018
00010982
09-11-2018
B
00011075
00011144
C
10-25-2018
12-05-2018
00011183
D
00011257
02-21-2019
E
00011tbd
G
09-16-2019