PLZ-5W 1
List of Messages
SCPI command: Command name in the short form.
*RST: "Yes" for commands that are affected by *RST.
R/W: “R” for query commands and “W” for set commands.
SOURce subsystem
SCPI command
Respo
nse
*RST Description
R/W
Program header
Parameter
ARB:APPL
Applies the I-V characteristics map and reflects it in the
output.
W
ARB:CLE
Returns the I-V characteristics map to its default set-
ting.
W
ARB:COUN numeric NR1 Yes
Sets the number of data points of the I-V characteris-
tics map.
R/W
ARB:DATA numeric NRf Yes Sets the I-V characteristics map in block data format.
R/W
ARB:MAP numeric NR3 Yes
Sets the voltage and current values for the specified
parts of the I-V characteristics map.
R/W
ARB:MAP:LIST? NR3
Queries the entire contents of the I-V characteristics
map.
R
COND numeric NR3 Yes Sets the conductance in CR mode.
R/W
COND:EXT:FCON bool bool Yes
Enables/disables external control of CR mode.
(CURR:EXT:FCON alias)
R/W
COND:PULS:LEV numeric NR3 Yes Sets the switching conductance.
R/W
COND:PULS:FREQ numeric NR3 Yes
Sets the switching frequency.
(CURR:PULS:FREQ alias)
R/W
COND:PULS:DCYC numeric NR3 Yes
Sets the switching duty cycle.
(CURR:PULS:DCYC alias)
R/W
COND:RANG LOW|MED|HIGH char Yes
Sets the CR mode range. (CURR:RANG alias)
R/W
COND:RESP NORM|FAST char Yes Sets the response speed in CR mode.
R/W
CURR numeric NR3 Yes Sets the current in CC mode.
R/W
CURR:EXT:ACON bool bool Yes
Enables/disables external superimposed input in CC
mode.
R/W
CURR:EXT:FCON bool bool Yes Enables/disables external control in CC mode.
R/W
CURR:PROT numeric NR3 Yes
Sets the current value for overcurrent protection
(OCP).
R/W
CURR:PROT:STAT bool bool Yes
Sets whether to turn the load off when overcurrent pro-
tection (OCP) is activated.
R/W
CURR:PULS:LEV numeric NR3 Yes Sets the switching current.
R/W
CURR:PULS:FREQ numeric NR3 Yes Sets the switching frequency.
R/W
CURR:PULS:DCYC numeric NR3 Yes Sets the switching duty cycle.
R/W
CURR:RANG LOW|MED|HIGH char Yes Set the range of CC mode.
R/W
CURR:SLEW numeric NR3 Yes Sets the slew rate value.
R/W
CURR:SST numeric NR3 Yes Sets the current’s soft start time.
R/W
FUNC CC|CR|CV|CP|ARB char Yes Sets the operation mode.
R/W
FUNC:CVOP bool bool Yes Enables or disables +CV mode.
R/W
POW numeric NR3 Yes Sets the power of CP mode.
R/W
POW:EXT:FCON bool bool Yes
Enables/disables external control in CP mode.
(CURR:EXT:FCON alias)
R/W
POW:PROT numeric NR3 Yes Sets the power of overpower protection (OPP).
R/W
POW:PROT:STAT bool bool Yes
Sets whether to turn the load off when overpower pro-
tection (OPP) is activated.
R/W
POW:RANG LOW|MED|HIGH char Yes
Sets the CP mode range. (CURR:RANG alias)
R/W
2 PLZ-5W
PROG string string Yes Selects or deselects a programs.
R/W
PROG:CRE string Creates a new program of the specified name.
W
PROG:CURR:PROT numeric NR3
Sets the current value for overcurrent protection (OCP)
in the selected program.
R/W
PROG:CURR:PROT:STAT bool bool
Sets the operation when overcurrent protection (OCP)
is activated in the selected program.
R/W
PROG:CVOP:LEV numeric NR3
Sets the voltage at +CV (addition of CV mode)
application in the selected program.
R/W
PROG:DEL string Deletes the program of the specified name.
W
PROG:LIST? string Retrieves the list of registered programs.
R
PROG:LOOP
numeric NR1 Sets the number of times that the selected program will
repeat.
R/W
INF char
PROG:POW:PROT numeric NR3
Sets the power of overpower protection (OPP) in the
selected program.
R/W
PROG:POW:PROT:STAT bool bool
Sets the operation to be performed when overpower
protection (OPP) is activated in the selected program.
R/W
PROG:REN string Changes the name of the selected program.
W
PROG:SAVE Saves the selected program.
W
PROG:STEP<n>:DWEL numeric NR3
Sets the step execution time at the specified step of the
selected program.
R/W
PROG:STEP<n>:INP bool bool
Sets whether to turn the load on at the specified step of
the selected program.
R/W
PROG:STEP<n>:LEV numeric NR3
Sets the load at the specified step of the selected pro-
gram.
R/W
PROG:STEP<n>:OUTP bool bool
Sets whether to turn the load on at the specified step of
the selected program.
R/W
PROG:STEP<n>:SLEW numeric NR3
Sets the slew rate at the specified step of the selected
program.
R/W
PROG:STEP<n>:TRIG:GEN char char
Sets whether to perform trigger output at the specified
step of the selected program.
R/W
PROG:STEP<n>:TRIG:WAIT char char
Sets trigger wait at the specified step of the selected
program.
R/W
PROG:STEPS:COUN numeric NR1
Sets the number of steps to be registered to the
selected program.
R/W
PROG:VOLT:PROT:LOW numeric NR3
Sets the voltage of underpower protection (UVP) in the
selected program.
R/W
PROG:VOLT:PROT:STAT bool bool
Enables/disables undervoltage protection (UVP) in the
selected program.
R/W
VOLT numeric NR3 Yes Sets the voltage of CV mode.
R/W
VOLT:EXT:FCON bool bool Enables/disables external control in CV mode.
R/W
VOLT:PROT:LOW numeric NR3 Yes Sets the voltage of underpower protection (UVP).
R/W
VOLT:PROT:STAT bool bool Yes Enables/disables undervoltage protection (UVP).
R/W
VOLT:RANG LOW|HIGH char Yes Sets the range of CV mode.
R/W
VOLT:RESP NORM|FAST char Yes Sets the response speed in CV mode.
R/W
SCPI command
Respo
nse
*RST Description
R/W
Program header
Parameter