Calibration technology
WIKA data sheet CT 15.01
Page 1 of 12
WIKA data sheet CT 15.01
11/2017
for further approvals
see page 4
ProcessCalibrator model CPH6000
Applications
Calibration service companies and service industry
Maintenance facilities
Measurement and control laboratories
Quality assurance
Special features
Digital indicator with easily interchangeable reference
pressure sensors (sensor can also be attached externally)
Measuring ranges from -1 ... 6,000 bar (-14.5 ... 87,000 psi)
Accuracy: 0.025 % (incl. calibration certicate)
Calibration and pressure-switch test functions
Software and complete service cases (incl. pumps)
available
ProcessCalibrator
Model CPH6000
Data sheets showing similar products and accessories:
Test pumps, hydraulic; model CPP series; see data sheet CT 91.05
Test pump, pneumatic; model CPP30; see data sheet CT 91.06
Calibration software; WIKA-Cal; see data sheet CT 95.10
Description
Handling
For a solution that always matches the application, there
are many pressure sensors to choose from, with accuracies
of up to 0.025 % and measuring ranges up to 6,000 bar
(87,000 psi), which can be interchanged quickly and
without tools. In addition to being xed to the instrument,
the reference pressure sensor can, optionally, be used
externally via an approx. 1.2 m (4 ft) long connecting cable. If
the reference pressure sensor on the instrument is changed,
then the digital indicator will recognise the new measuring
range and so this saves the operator from having to congure
it via the menu.
Functionality
In the setup menu there are 3 operating modes to choose
from:
MEASURING, CALIBRATION and PRESSURE SWITCH-
TEST. In the CALIBRATION and PRESSURE SWITCH-
TEST modes, a menu assistant supports the user during
each operation and records, for example, the calibration
data from several calibrations or automatically calculates
the switch hysteresis. To power the test item and to read its
measurement signals, there are electrical inputs and outputs
which are protected from adverse conditions in the eld by
captive protection caps.
Software
For the evaluation and documentation of the calibration
data stored in the CPH6000, WIKA-Cal calibration software
is available. Using this software, the data is automatically
transferred into a printable calibration certicate.
Furthermore, WIKA-Cal also oers, over and above
PC-supported calibration, the management of the calibration
and instrument data in an SQL database. For data transfer,
an RS-232 and a USB interface are available.
Page 2 of 12 WIKA data sheet CT 15.01 ∙ 11/2017
Specications
ProcessCalibrator model CPH6000
Sensor technology 1 reference pressure sensor (interchangeable without tools)
1)
;
Option: external operation via 1.2 m (4 ft) cable
Measuring range
Gauge bar -0.25 ... +0.25 -0.4 ... +0.4 -0.6 ... +0.6 -1 ... 0 -1 ... 1.5
-1 ... 2.5 -1 ... 3 -1 ... 5 -1 ... 9 -1 ... 15
-1 ... 24 -1 ... 39 0 ... 0.25 0 ... 0.4 0 ... 0.6
0 ... 1 0 ... 1.6 0 ... 2.5 0 ... 4 0 ... 6
0 ... 10 0 ... 16 0 ... 25 0 ... 40 0 ... 60
0 ... 100 0 ... 160 0 ... 250 0 ... 400 0 ... 600
0 ... 1,000
psi -4 ... +4 -6 ... +6 -10 ... +10 -14.5 ... 0 -14.5 ... 22
-14.5 ... 36 -14.5 ... 44 -14.5 ... 73 -14.5 ... 130 -14.5 ... 220
-14.5 ... 350 -14.5 ... 565 0 ... 4 0 ... 6 0 ... 10
0 ... 14,5 0 ... 25 0 ... 40 0 ... 60 0 ... 90
0 ... 145 0 ... 250 0 ... 360 0 ... 580 0 ... 870
0 ... 1,450 0 ... 2,320 0 ... 3,630 0 ... 5,800 0 ... 8,700
0 ... 14,500
Accuracy of the measuring chain 0.025 % FS
2)
Gauge bar 0 ... 1,600 0 ... 2,500 0 ... 4,000 0 ... 5,000 0 ... 6,000
psi 0 ... 23,200 0 ... 36,200 0 ... 58,000 0 ... 72,500 0 ... 87,000
Accuracy of the measuring chain 0.1 % FS
2)
Absolute pressure bar abs. -0.8 ... 1.2 0 ... 0.25 0 ... 0.4 0 ... 0.6 0 ... 1
0 ... 1.6 0 ... 2.5 0 ... 4 0 ... 6 0 ... 10
0 ... 16 0 ... 25
psi abs. -11.6 ... 17.4 0 ... 4 0 ... 6 0 ... 8 0 ... 14.5
0 ... 23 0 ... 36 0 ... 60 0 ... 90 0 ... 145
0 ... 230 0 ... 360
Accuracy of the measuring chain 0.025 % FS
2)
Overpressure limit
(dependent on the measuring
range)
3 times; < 25 bar
2 times; > 25 bar ... ≤ 600 bar
1,5 times; > 600 bar ... ≤ 2,500 bar
1,2 times; > 2,500 bar
3 times; < 360 psi
2 times; > 360 psi ... ≤ 8,700 psi
1,5 times; > 8,700 psi ... ≤ 36,260 psi
1,2 times; > 36,260 psi
Types of pressure Relative pressure, {absolute pressure from 0 ... 25 bar abs. (0 ... 360 psi abs.) and vacuum from
-1 ... 39 bar (-14.5 ... 550 psi)}
Sensor compatibility Compatible with model CPT6000 reference pressure sensors
{} Items in curved brackets are optional extras for an additional price.
1) Up to 10 reference pressure sensors can be supported on each instrument (up to 10 calibration data sets).
2) Calibrated at 23 °C (74 °F) and in vertical mounting position, pressure connection facing downwards.
Complete test and service cases
For maintenance and service applications, various service
case systems are available. These include service cases
with or without pressure generation, charger, test-cable set,
adapter, etc.
Certied accuracy
For each reference pressure sensor, the accuracy is certied
by a factory calibration certicate which accompanies the
instrument. On request, we can provide a DKD/DAkkS
calibration certicate for this instrument.