General
Specications
<<Contents>> <<Index>>
GS 04L41B01-01E
Daqstation
DX1000
OVERVIEW
The DX1000 is a DAQSTATION that displays real-
time measured data on a color LCD and saves data
on a CompactFlash memory card (CF card). It can
be hooked up to network via Ethernet, which enables
to inform by E-mail and to monitor on Web site as
well as to transfer les by using FTP. Also, it can
communicate with Modbus/RTU or Modbus/TCP.
It comes with a two, four, six-channel or twelve-
channel model. As the input signal, a DC voltage,
thermocouple, resistance temperature detector,
or contact signal can be set to each channel. The
data saved on a CF card can be converted by data
conversion software to Lotus 1-2-3, Excel, or ASCII
format le, facilitating processing on a PC. Not only
this, the Viewer software allows a PC to display
waveforms on its screen and to print out waveforms.
STANDARD SPECIFICATIONS
General Specications
Construction
Mounting: Flush panel mounting (on a vertical plane)
Mounting may be inclined downward up to
30 degrees from a horizontal plane.
Allowable panel thickness:
2 to 26 mm
Material: Case: drawn steel
Bezel: polycarbonate
Displaylter:polycarbonate
Case color:
Case: Grayish blue green
(Munsell 2.0B 5.0/1.7 or equivalent)
Bezel: Charcoal grey light
(Munsell 10B 3.6/0.3 or equivalent)
Front panel:
Water and dust-proof*
(based on IEC529-IP65 and NEMA
No.250 TYPE4 for indoor locations
(except external icing test))
*Except for side-by-side mounting.
Dimensions:
144 (W) × 144 (H) × 224.1 (D) mm
144 (W) × 144 (H) × 228.5 (D)* mm
*Incaseof/H2or/PM1optionisspecied.
Weight: approx. 2.9 kg*
*without optional features
Input
Number of inputs:
DX1002: two channels
DX1004: four channels
DX1006: six channels
DX1012: twelve channels
Measurement interval:
DX1002, DX1004:
125 ms, 250 ms, 25 ms (fast sampling
mode*)
DX1006, DX1012:
1 s (Not available when A/D integration
time is set to 100 ms), 2 s, 5 s, 125 ms
(fast sampling mode*)
*A/Dintegrationtimeisxedto1.67msin
case of fast sampling mode.
Inputs: DCV (DC voltage), TC (thermocouple),
RTD (resistance temperature detector),
DI (digital input for event recording), DCA
(DC current with external shunt resistor
attached)
GS 04L41B01-01E
©Copyright November. 2005
1st Edition Nov. 2005(KP)
18th Edition Feb. 2017(YK)
2
All Rights Reserved. Copyright © 2005, Yokogawa Electric Corporation
<<Contents>> <<Index>>
GS 04L41B01-01E Feb. 3, 2017-00
Input
type
Range Measuring range
DCV
20 mV –20.000 to 20.000 mV
60 mV –60.00 to 60.00 mV
200 mV –200.00 to 200.00 mV
2 V –2.0000 to 2.0000 V
6 V –6.000 to 6.000 V
1-5V –0.800 to 5.200 V
20 V –20.000 to 20.000 V
50 V –50.00 to 50.00 V
TC
R*
1
0.0 to 1760.0°C 32 to 3200°F
S*
1
0.0 to 1760.0°C 32 to 3200°F
B*
1
0.0 to 1820.0°C 32 to 3308°F
K*
1
–200.0 to 1370.0°C –328 to 2498°F
E*
1
–200.0 to 800.0°C –328.0 to 1472.0°F
J*
1
–200.0 to 1100.0°C –328.0 to 2012.0°F
T*
1
–200.0 to 400.0°C –328.0 to 752.0°F
N*
1
–270.0 to 1300.0°C –454 to 2372°F
W*
2
0.0 to 2315.0°C 32 to 4199°F
L*
3
–200.0 to 900.0°C –328.0 to 1652.0°F
U*
3
–200.0 to 400.0°C –328.0 to 752.0°F
WRe*
4
0.0 to 2400.0°C 32 to 4352°F
RTD
Pt100*
5
–200.0 to 600.0°C –328.0 to 1112.0°F
JPt100*
5
–200.0 to 550.0°C –328.0 to 1022.0°F
DI
DCV
input
(TTL)
OFF : less than 2.4 V
ON : more than 2.4 V
Contact
input
Contact ON/OFF
*1 R, S, B, K, E, J, T, N: IEC 60584-1, DIN EN
60584-1,JIS C 1602
*2 W: W-5% Re/W-26% Re (Hoskins Mfg. Co.),
ASTM E988-96
(Type C equivalent of OMEGA Engineering Inc.)
*3 L: Fe-CuNi, DIN43710, U: Cu-CuNi, DIN43710
*4 WRe: W-3%Re/W-25%Re (Hoskins Mfg. Co.),
ASTM E988-96
(Type D equivalent of OMEGA Engineering Inc.)
*5 Pt100: JIS C 1604, IEC 60751, DIN EN 60751
JPt100: JIS C 1604, JIS C 1606
Measuring current: i = 1mA
A/D integration time:
20 ms (50 Hz), 16.7 ms (60 Hz), 100ms
(50/60Hz for DX1006/1012), or AUTO
selectable (automatic selection by
detection of power supply frequency)
A/Dintegrationtimeisxedto1.67ms
(600Hz) in case of fast sampling mode.
Thermocouple burnout:
Burnout upscale/downscale function can
be switched on/off (for each channel).
Burnout upscale/downscale selectable
Normal:Lessthan2kΩ,Burnout:More
than100kΩ
Detection current: approx. 10 µA
1-5V range burnout:
Burnout upscale/downscale function can
be switched on/off (for each channel).
Burnout upscale/downscale selectable
Upscale burnout: More than +10% of
conguredspan
Downscale burnout: Less than -5% of
conguredspan
Moving average:
Moving average on/off selectable for
each channel
Moving average cycles 2 to 400 selectable
Calculation:
Differential computation:
Between any two channels
Available for DCV, TC, RTD and DI ranges.
Linear scaling:
Available for DCV, TC, RTD and DI
ranges.
Scaling limits: -30000 to 30000
Decimal point: user-selectable
Engineeringunit:user-denable,upto6
characters
Over value: Exceeds ± 5% of scaling
limits (on/off selectable)
Square root:
Available for DCV range.
Scaling limits: -30000 to 30000
Decimal point: user-selectable
Engineeringunit:user-denable,upto6
characters
Low level cut off: 0.0 to 5.0% of display
span
Over value: Exceeds ± 5% of scaling
limits (on/off selectable)
1-5VDC scaling:
Available for 1-5VDC range.
Scaling limits: -30000 to 30000
Display span limit: 0.800 to 5.200
Decimal point: user-selectable
Engineeringunit:user-denable,upto6
characters
Low level cut off: Fixed to lower span limit
Over value: Exceeds ± 5% of scaling
limits (on/off selectable)
Display
Display unit:
5.7-inch TFT color LCD (VGA, 320 × 240
pixels)
Note) In the part of crystal display, there are some pixels
that can’t always turn on or off. Please understand
that the brightness of screen looks uneven
because of characteristics of crystal display, but it
is not out of order.
Display group:
Each measurement channel and computation
channel can be assigned to display group of the
trend, digital and bargraph display.
Number of display: 10 groups
Number of assignable channels for one
group: 6 channels
Display color:
Trend/Bargraph:
Selectable from 24 colors
Background:
White or black selectable
Trend display:
Trend display type:
Vertical, horizontal, landscape,
horizontal or split selectable
Number of indication channels:
6 channels per display (maximum)
Number of display:
10 displays (10 groups)
Line width:
1, 2, and 3 pixels selectable