MIHW3000 SERIES
DC-DC CONVERTER 6W, Reinforced Insulation, Medical Safety
E-mail:sales@minmax.com.tw Tel:886-6-2923150
2018/10/31 REV:18 Page 1 of 5
®
FEATURES
Industrial Standard DIP-24 Package
Wide 2:1 Input Voltage Range
Fully Regulated Output Voltage
I/O Isolation 4000VAC with Reinforced Insulation, rated for 1000Vrms
Working Voltage
Low Leakage Current < 2µA
Operating Ambient Temp. Range -40 to +75
Under-Voltage, Overload and Short Circuit Protection
Conducted EMI EN 55011/22 Class A & FCC Level A Approved
Medical EMC Standard with 4
th
Edition of EMI EN 55011 and EMS EN 60601-1-2
Approved
Medical Safety with 1xMOPP & 2xMOOP per 3
rd
Edition of IEC/EN 60601-1 &
ANSI/AAMI ES60601-1 Approved
UL/cUL/IEC/EN 60950-1 Safety Approval & CE Marking
PRODUCT OVERVIEW
The MINMAX MIHW3000 series is a range of high performance DC-DC converter modules with a reinforced insulation system .The I/O isolation voltage is
specified for 4000VAC with reinforced insulation, which rated for 1000Vrms working voltage.The product comes in a small DIP-24 package. All 12 models
features wide 2:1 input voltage range and fully regulated output voltage.
The MIHW3000 DC-DC converters offer an economical solution for demanding applications in industrial and medical instrumentation requesting a certified
high supplementary or reinforced insulation system to comply with relative industrial or medical safety standards.
Model Selection Guide
Model
Number
Input
Voltage
(Range)
Output
Voltage
Input Current
Reflected
Ripple
Current
Max. capacitive
Load
Efficiency
(typ.)
Max.
Min.
@Max. Load
@No Load
@Max. Load
VDC
VDC
mA
mA
mA(typ.)
mA(typ.)
mA (typ.)
μF
%
MIHW3022
12
(9 ~ 18)
5
1000
200
570
30
60
1000
75
MIHW3023
12
500
100
641
470
78
MIHW3026
±12
±250
±50
641
220#
78
MIHW3027
±15
±200
±40
641
220#
78
MIHW3032
24
(18 ~ 36)
5
1000
200
278
20
30
1000
77
MIHW3033
12
500
100
313
470
80
MIHW3036
±12
±250
±50
313
220#
80
MIHW3037
±15
±200
±40
313
220#
80
MIHW3042
48
(36 ~ 75)
5
1000
200
139
10
15
1000
77
MIHW3043
12
500
100
156
470
80
MIHW3046
±12
±250
±50
156
220#
80
MIHW3047
±15
±200
±40
156
220#
80
# For each output
MIHW3000 SERIES
DC-DC CONVERTER 6W, Reinforced Insulation, Medical Safety
E-mail:sales@minmax.com.tw Tel:886-6-2923150
2018/10/31 REV:18 Page 2 of 5
®
Input Specifications
Parameter
Model
Min.
Typ.
Max.
Unit
Input Surge Voltage (1 sec. max.)
12V Input Models
-0.7
---
25
VDC
24V Input Models
-0.7
---
50
48V Input Models
-0.7
---
100
Start-Up Threshold Voltage
12V Input Models
7
8
9
24V Input Models
13
15
18
48V Input Models
30
33
36
Under Voltage Shutdown
12V Input Models
---
---
8.5
24V Input Models
---
---
16
48V Input Models
---
---
34
Short Circuit Input Power
All Models
---
---
3000
mW
Input Filter
Internal Pi Type
Conducted EMI
Compliance to EN 55011/22,class A and FCC part 15,class A
Output Specifications
Parameter
Conditions / Model
Min.
Typ.
Max.
Unit
Output Voltage Setting Accuracy
---
---
±1.0
%Vnom.
Output Voltage Balance
Dual Output, Balanced Loads
---
±0.5
±2.0
%
Line Regulation
Vin=Min. to Max. @Full Load
---
±0.3
±0.5
%
Load Regulation
Io=25% to 100%
---
±0.5
±1.0
%
Ripple & Noise
0-20 MHz Bandwidth
5V Output Models
---
75
100
mV
P-P
Other Output Models
---
100
150
mV
P-P
Transient Recovery Time
25% Load Step Change
---
300
500
μsec
Transient Response Deviation
---
±3
±6
%
Temperature Coefficient
---
±0.02
±0.05
%/
Over Load Protection
Foldback
120
150
---
%
Short Circuit Protection
Continuous, Automatic Recovery
Isolation, Safety Standards
Parameter
Conditions
Min.
Typ.
Max.
Unit
I/O Isolation Voltage
60 Seconds
Reinforced insulation, rated for 1000Vrms working voltage
4000
---
---
VACrms
Leakage Current
240VAC, 60Hz
---
---
2
μA
I/O Isolation Resistance
500 VDC
10
---
---
GΩ
I/O Isolation Capacitance
100kHz, 1V
---
7
13
pF
Safety Standards
UL/cUL 60950-1, CSA C22.2 No. 60950-1
ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1
IEC/EN 60950-1, IEC/EN 60601-1 3
rd
Edition 1xMOPP & 2xMOOP
Safety Approvals
UL/cUL 60950-1 recognition (UL certificate), IEC/EN 60950-1 (CB-report)
ANSI/AAMI ES60601-1 1xMOPP & 2xMOOP recognition (UL certificate), IEC/EN 60601-1 3
rd
Edition (CB-report)
General Specifications
Parameter
Conditions
Min.
Typ.
Max.
Unit
Switching Frequency
---
150
---
kHz
MTBF(calculated)
MIL-HDBK-217F@25, Ground Benign
1,000,000
---
---
Hours