MIHW1000 SERIES
DC-DC CONVERTER 3W, Reinforced Insulation, Medical Safety
E-mail:sales@minmax.com.tw Tel:886-6-2923150
2018/10/31 REV:14 Page 1 of 4
®
FEATURES
Industrial Standard DIP-24 Package
Wide 2:1 Input Voltage Range
Fully Regulated Output Voltage
I/O Isolation 4000VAC with Reinforced Insulation, rated for 1000Vrms
Working Voltage
Low Leakage Current < 2µA
Operating Ambient Temp. Range -40 to +85
Under-Voltage, Overload and Short Circuit Protection
Conducted EMI EN 55011/22 Class A & FCC Level A Approved
Medical EMC Standard with 4
th
Edition of EMI EN 55011 and EMS EN 60601-1-2
Approved
Medical Safety with 1xMOPP & 2xMOOP per 3
rd
Edition of IEC/EN 60601-1 &
ANSI/AAMI ES60601-1 Approved
UL/cUL/IEC/EN 60950-1 Safety Approval & CE Marking
PRODUCT OVERVIEW
The MINMAX MIHW1000 series is a range of high performance DC-DC converter modules with a reinforced insulation system .The I/O isolation voltage is
specified for 4000VAC with reinforced insulation, which rated for 1000Vrms working voltage.The product comes in a small DIP-24 package. All 20 models
features wide 2:1 input voltage range and fully regulated output voltage.
The MIHW1000 DC-DC converters offer an economical solution for demanding applications in industrial and medical instrumentation requesting a certified
supplementary or reinforced insulation system to comply with relative industrial or medical safety standards.
Model Selection Guide
Model
Number
Input
Voltage
(Range)
Output
Voltage
Input Current
Reflected
Ripple
Current
Max. capacitive
Load
Efficiency
(typ.)
Max.
Min.
@Max. Load
@No Load
@Max. Load
VDC
VDC
mA
mA
mA(typ.)
mA(typ.)
mA (typ.)
μF
%
MIHW1002
5
(4.5 ~ 9)
5
600
90
857
40
60
1000
70
MIHW1003
12
250
37.5
800
470
75
MIHW1008
24
125
18.8
800
470
76
MIHW1006
±12
±125
±18.8
800
220#
75
MIHW1007
±15
±100
±15
800
220#
75
MIHW1012
12
(9 ~ 18)
5
600
90
338
30
30
1000
74
MIHW1013
12
250
37.5
313
470
80
MIHW1018
24
125
18.8
313
470
81
MIHW1016
±12
±125
±18.8
313
220#
80
MIHW1017
±15
±100
±15
313
220#
80
MIHW1022
24
(18 ~ 36)
5
600
90
160
20
15
1000
78
MIHW1023
12
250
37.5
151
470
83
MIHW1028
24
125
18.8
151
470
84
MIHW1026
±12
±125
±18.8
151
220#
83
MIHW1027
±15
±100
±15
151
220#
83
MIHW1032
48
(36 ~ 75)
5
600
90
80
10
10
1000
78
MIHW1033
12
250
37.5
75
470
83
MIHW1038
24
125
18.8
75
470
84
MIHW1036
±12
±125
±18.8
75
220#
83
MIHW1037
±15
±100
±15
75
220#
83
# For each output
MIHW1000 SERIES
DC-DC CONVERTER 3W, Reinforced Insulation, Medical Safety
E-mail:sales@minmax.com.tw Tel:886-6-2923150
2018/10/31 REV:14 Page 2 of 4
®
Input Specifications
Parameter
Model
Min.
Typ.
Max.
Unit
Input Surge Voltage (1 sec. max.)
5V Input Models
-0.7
---
11
VDC
12V Input Models
-0.7
---
25
24V Input Models
-0.7
---
50
48V Input Models
-0.7
---
100
Start-Up Threshold Voltage
5V Input Models
3.7
4
4.5
12V Input Models
8
8.5
9
24V Input Models
15
17
18
48V Input Models
30
33
36
Under Voltage Shutdown
5V Input Models
---
---
4
12V Input Models
---
---
8.5
24V Input Models
---
---
17
48V Input Models
---
---
34
Short Circuit Input Power
All Models
---
---
2000
mW
Input Filter
Internal Pi Type
Conducted EMI
Compliance to EN 55011/22,class A and FCC part 15,class A
Output Specifications
Parameter
Conditions / Model
Min.
Typ.
Max.
Unit
Output Voltage Setting Accuracy
---
---
±1.0
%Vnom.
Output Voltage Balance
Dual Output, Balanced Loads
---
±0.5
±2.0
%
Line Regulation
Vin=Min. to Max.
---
±0.3
±0.5
%
Load Regulation
Io=25% to 100%
---
±0.5
±1.0
%
Ripple & Noise
0-20 MHz Bandwidth
5V Output Models
---
75
100
mV
P-P
Other Output Models
---
100
150
mV
P-P
Transient Recovery Time
25% Load Step Change
---
150
500
μsec
Transient Response Deviation
---
±3
±6
%
Temperature Coefficient
---
±0.02
±0.05
%/
Over Load Protection
Foldback
120
150
---
%
Short Circuit Protection
Continuous, Automatic Recovery
Isolation, Safety Standards
Parameter
Conditions
Min.
Typ.
Max.
Unit
I/O Isolation Voltage
60 Seconds
Reinforced insulation, rated for 1000Vrms working voltage
4000
---
---
VACrms
Leakage Current
240VAC, 60Hz
---
---
2
μA
I/O Isolation Resistance
500 VDC
10
---
---
GΩ
I/O Isolation Capacitance
100kHz, 1V
---
7
13
pF
Safety Standards
UL/cUL 60950-1, CSA C22.2 No. 60950-1
ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No. 60601-1
IEC/EN 60950-1, IEC/EN 60601-1 3
rd
Edition 1xMOPP & 2xMOOP
Safety Approvals
UL/cUL 60950-1 recognition (UL certificate), IEC/EN 60950-1 (CB-report)
ANSI/AAMI ES60601-1 1xMOPP & 2xMOOP recognition (UL certificate), IEC/EN 60601-1 3
rd
Edition (CB-report)
General Specifications
Parameter
Conditions
Min.
Typ.
Max.
Unit
Switching Frequency
---
150
---
kHz
MTBF(calculated)
MIL-HDBK-217F@25, Ground Benign
1,000,000
---
---
Hours