General
Specications
<<Contents>> <<Index>>
GS 04L41B01-01E
Daqstation
DX1000
OVERVIEW
The DX1000 is a DAQSTATION that displays real-
time measured data on a color LCD and saves data
on a CompactFlash memory card (CF card). It can
be hooked up to network via Ethernet, which enables
to inform by E-mail and to monitor on Web site as
well as to transfer les by using FTP. Also, it can
communicate with Modbus/RTU or Modbus/TCP.
It comes with a two, four, six-channel or twelve-
channel model. As the input signal, a DC voltage,
thermocouple, resistance temperature detector,
or contact signal can be set to each channel. The
data saved on a CF card can be converted by data
conversion software to Lotus 1-2-3, Excel, or ASCII
format le, facilitating processing on a PC. Not only
this, the Viewer software allows a PC to display
waveforms on its screen and to print out waveforms.
STANDARD SPECIFICATIONS
General Specications
Construction
Mounting: Flush panel mounting (on a vertical plane)
Mounting may be inclined downward up to
30 degrees from a horizontal plane.
Allowable panel thickness:
2 to 26 mm
Material: Case: drawn steel
Bezel: polycarbonate
Displaylter:polycarbonate
Case color:
Case: Grayish blue green
(Munsell 2.0B 5.0/1.7 or equivalent)
Bezel: Charcoal grey light
(Munsell 10B 3.6/0.3 or equivalent)
Front panel:
Water and dust-proof*
(based on IEC529-IP65 and NEMA
No.250 TYPE4 for indoor locations
(except external icing test))
*Except for side-by-side mounting.
Dimensions:
144 (W) × 144 (H) × 224.1 (D) mm
144 (W) × 144 (H) × 228.5 (D)* mm
*Incaseof/H2or/PM1optionisspecied.
Weight: approx. 2.9 kg*
*without optional features
Input
Number of inputs:
DX1002: two channels
DX1004: four channels
DX1006: six channels
DX1012: twelve channels
Measurement interval:
DX1002, DX1004:
125 ms, 250 ms, 25 ms (fast sampling
mode*)
DX1006, DX1012:
1 s (Not available when A/D integration
time is set to 100 ms), 2 s, 5 s, 125 ms
(fast sampling mode*)
*A/Dintegrationtimeisxedto1.67msin
case of fast sampling mode.
Inputs: DCV (DC voltage), TC (thermocouple),
RTD (resistance temperature detector),
DI (digital input for event recording), DCA
(DC current with external shunt resistor
attached)
GS 04L41B01-01E
©Copyright November. 2005
1st Edition Nov. 2005(KP)
19th Edition Apr. 2019(YK)
2
All Rights Reserved. Copyright © 2005, Yokogawa Electric Corporation
<<Contents>> <<Index>>
GS 04L41B01-01E Apr. 24, 2019-00
Input
type
Range Measuring range
DCV
20 mV –20.000 to 20.000 mV
60 mV –60.00 to 60.00 mV
200 mV –200.00 to 200.00 mV
2 V –2.0000 to 2.0000 V
6 V –6.000 to 6.000 V
1-5V –0.800 to 5.200 V
20 V –20.000 to 20.000 V
50 V –50.00 to 50.00 V
TC
R*
1
0.0 to 1760.0°C 32 to 3200°F
S*
1
0.0 to 1760.0°C 32 to 3200°F
B*
1
0.0 to 1820.0°C 32 to 3308°F
K*
1
–200.0 to 1370.0°C –328 to 2498°F
E*
1
–200.0 to 800.0°C –328.0 to 1472.0°F
J*
1
–200.0 to 1100.0°C –328.0 to 2012.0°F
T*
1
–200.0 to 400.0°C –328.0 to 752.0°F
N*
1
–270.0 to 1300.0°C –454 to 2372°F
W*
2
0.0 to 2315.0°C 32 to 4199°F
L*
3
–200.0 to 900.0°C –328.0 to 1652.0°F
U*
3
–200.0 to 400.0°C –328.0 to 752.0°F
WRe*
4
0.0 to 2400.0°C 32 to 4352°F
RTD
Pt100*
5
–200.0 to 600.0°C –328.0 to 1112.0°F
JPt100*
5
–200.0 to 550.0°C –328.0 to 1022.0°F
DI
DCV
input
(TTL)
OFF : less than 2.4 V
ON : more than 2.4 V
Contact
input
Contact ON/OFF
*1 R, S, B, K, E, J, T, N: IEC 60584-1, DIN EN
60584-1,JIS C 1602
*2 W: W-5% Re/W-26% Re (Hoskins Mfg. Co.),
ASTM E988-96
(Type C equivalent of OMEGA Engineering Inc.)
*3 L: Fe-CuNi, DIN43710, U: Cu-CuNi, DIN43710
*4 WRe: W-3%Re/W-25%Re (Hoskins Mfg. Co.),
ASTM E988-96
(Type D equivalent of OMEGA Engineering Inc.)
*5 Pt100: JIS C 1604, IEC 60751, DIN EN 60751
JPt100: JIS C 1604, JIS C 1606
Measuring current: i = 1mA
A/D integration time:
20 ms (50 Hz), 16.7 ms (60 Hz), 100ms
(50/60Hz for DX1006/1012), or AUTO
selectable (automatic selection by
detection of power supply frequency)
A/Dintegrationtimeisxedto1.67ms
(600Hz) in case of fast sampling mode.
Thermocouple burnout:
Burnout upscale/downscale function can
beswitchedon/o(foreachchannel).
Burnout upscale/downscale selectable
Normal:Lessthan2kΩ,Burnout:More
than100kΩ
Detection current: approx. 10 µA
1-5V range burnout:
Burnout upscale/downscale function can
beswitchedon/o(foreachchannel).
Burnout upscale/downscale selectable
Upscale burnout: More than +10% of
conguredspan
Downscale burnout: Less than -5% of
conguredspan
Moving average:
Movingaverageon/oselectablefor
each channel
Moving average cycles 2 to 400 selectable
Calculation:
Dierentialcomputation:
Between any two channels
Available for DCV, TC, RTD and DI ranges.
Linear scaling:
Available for DCV, TC, RTD and DI
ranges.
Scaling limits: -30000 to 30000
Decimal point: user-selectable
Engineeringunit:user-denable,upto6
characters
Over value: Exceeds ± 5% of scaling
limits(on/oselectable)
Square root:
Available for DCV range.
Scaling limits: -30000 to 30000
Decimal point: user-selectable
Engineeringunit:user-denable,upto6
characters
Lowlevelcuto:0.0to5.0%ofdisplay
span
Over value: Exceeds ± 5% of scaling
limits(on/oselectable)
1-5VDC scaling:
Available for 1-5VDC range.
Scaling limits: -30000 to 30000
Display span limit: 0.800 to 5.200
Decimal point: user-selectable
Engineeringunit:user-denable,upto6
characters
Lowlevelcuto:Fixedtolowerspanlimit
Over value: Exceeds ± 5% of scaling
limits(on/oselectable)
Display
Display unit:
5.7-inch TFT color LCD (VGA, 320 × 240
pixels)
Note) In the part of crystal display, there are some pixels
thatcan’talwaysturnonoro.Pleaseunderstand
that the brightness of screen looks uneven
because of characteristics of crystal display, but it
is not out of order.
Display group:
Each measurement channel and computation
channel can be assigned to display group of the
trend, digital and bargraph display.
Number of display: 10 groups
Number of assignable channels for one
group: 6 channels
Display color:
Trend/Bargraph:
Selectable from 24 colors
Background:
White or black selectable
Trend display:
Trend display type:
Vertical, horizontal, landscape,
horizontal or split selectable
Number of indication channels:
6 channels per display (maximum)
Number of display:
10 displays (10 groups)
Line width:
1, 2, and 3 pixels selectable