dwyer-inst.com
DWYER INSTRUMENTS, INC.
T: 219-879-8000 | F: 219-872-9057 | info@dwyermail.com
800-872-9141
SERIES EDA | MERCOID
®
ELECTRONIC PRESSURE CONTROLLER
FEATURES/BENEFITS
Versatile compact package that can replace a separate gage, two
switches, and a transmitter in a system saving money, installation time,
and panel space
Fully programmable to meet simple or complex application needs
APPLICATIONS
Process control
Compressor control
DESCRIPTION
Series EDA Electronic Pressure Control is an extremely versatile compact package that
can replace a separate gage, two switches, and a transmitter in a system saving money,
installation time, and panel space. The EDA incorporates two SPDT relays that
have the on and off points fully adjustable over the range for control or alarm use. Front
face has LED indicators for switch status and a large backlight two-line display showing
process value and indication units. Programming is easy with simple menu structure, two-line
display, and external programming buttons. Weatherproof housing is ideal for a wide variety
of applications with panel mount, ush mount, or pipe mount ability
SPECIFICATIONS
SWITCH SPECIFICATIONS
TRANSMITTER SPECIFICATIONS
Service Compatible liquids and gases.
Wetted Materials 316L SS.
Housing Glass lled plastic.
Accuracy ±1% of FS including linearity, hysteresis, and repeatability (indicator and transmitter).
Stability < ±2% of FS per year.
Pressure Limits 1.5 x range.
Temperature Limits Ambient: 20 to 140°F (-6.6 to 60°C); Process: 0 to 176°F (-18 to 80°C).
Compensated Temperature Limits 32 to 122°F (0 to 50°C).
Thermal Effect ±0.05% of F.S./°F.
Display 4-digit backlit LCD (Digits: 0.60˝ H x 0.33˝ W).
Power Requirements 12 to 30 VDC/AC.
Power Consumption 2.5 watts.
Electrical Connections Removable terminal blocks with 1/2˝ female NPT conduit connections.
Enclosure Rating Meets NEMA 4X (IP65).
Warm Up Time <10 seconds.
Mounting Orientation Any position.
Weight 1.18 lbs (535 g).
Agency Approvals CE, UL.
Switch Type 2 SPDT relays.
Electrical Rating 5 A @ 120/240 VAC, 1 A @ 30 VDC.
Repeatability ±1% of FS (switching only).
Set Points Adjustable 0-100% of FS.
Switch Indication External LED for each relay on the front panel.
Switch Reset Manual or Automatic.
Output Signal 4 to 20 mA, 1 to 6 VDC, 1 to 5 VDC, or 0 to 10 VDC (direct or reverse output selection).
Minimum Excitation 14 VDC.
Zero and Span Adjustments Menu scalable within the range.
®
dwyer-inst.com
DWYER INSTRUMENTS, INC.
T: 219-879-8000 | F: 219-872-9057 | info@dwyermail.com
800-872-9141
Important Notice: Dwyer Instruments, Inc. reserves the right to make changes to or discontinue any product or service identied in this publication without notice. Dwyer advises its customers to obtain the
latest version of the relevant information to verify, before placing any orders, that the information being relied upon is current.
DS-EDA | 3/18
DIMENSIONS WIRING DIAGRAM
HOW TO ORDER
SERIES
EDA: Electronic pressure controller
HOUSING
W: Weatherproof
PROCESS CONNECTION
-N1: 1/4˝ male NPT
CONDUIT CONNECTION
E1: Two 1/2˝ NPT
OPTIONS
-AT: Aluminum adhesive tag
-NIST: NIST certicate
-P: 1/2 NPSM conduit plug
-23444: Special cleaning
TRANSMITTER OUTPUT
T0: None
T1: 4 to 20 mA
T2: 1 to 5 VDC
T3: 0 to 5 VDC
T4: 1 to 6 VDC
T5: 0 to 10 VDC
RANGE
-02: 0 to 20 psig
-03: 0 to 60 psig
-04: 0 to 100 psig
-05: 0 to 150 psig
-06: 0 to 300 psig
-07: 0 to 600 psig
-08: 0 to 1000 psig
-09: 0 to 1500 psig
-10: 0 to 3000 psig
EDA E1W -02-N1 T3 -NIST
Use the bold characters from the chart below to construct a product code.
ACCESSORIES
Model Description
A-590
A-370
1/2˝ conduit plug, watertight.
Flush mount bracket used with: Photohelic
®
, 3000MR/MRS, Capsuhelic
®
, 631 B Transmitter, EDA. Bracket is then surface mounted. Steel with
gray hammertone epoxy nish
1/4 MALE NPT
WITH 3/4 [19.05]
WRENCHING HEX
9/32 [7.14]
Ø 5
[127.00]
Ø 4-17/32
[115.09]
1
[25.40]
3-1/16
[77.89]
27/32 [21.43]
1-59/64
[26.16]
2X 5/16
[7.93]
3-15/32
[88.11]
1-5/8
[41.28]
[40.88]
3X 1/8 [3.18]
1/2 FEMALE
NPT CONDUIT
ENTRY
SP1
RELAY
CONN7 CONN8
CONN9
CONN10
LOAD 1
SP2
RELAY
SP2 OR
ALARM
RELAY
5A
250 VAC FOR LOADS USING 120/240 VAC POWER
1A 32 VDC FOR LOADS USING 30 VDC POWER
375 mA 250 VAC
12 TO 30
VDC/AC POWER
LINE
INPUT
RECEIVER
VOLTAGE
OUTPUT
RECEIVER
LOAD 2
SP2 OR ALARM
RELAY
LINE
INPUT
4 TO 20mA
OUTPUT
NO
NONC
NC