General Information Regarding the CE Marking
General Information
Regarding the
CE Marking
General Information Regarding the CE Marking
All ROHDE & SCHWARZ measuring instruments comply
with the regulations of the EMC directive. The ROHDE &
SCHWARZ conformity evaluation is based on applicable
professional and product standards. In cases where
multiple limit values can be used, ROHDE & SCHWARZ
will apply the stricter standard of the two. With regard to
interference emission, the limit values for businesses and
commercial operations including the small business sector
are applied (Class 1B). For the susceptibility to interference
emissions, the applicable limit values for the industrial sec-
tor are used. The signal and data cables that must be con-
nected to the device during operation will necessarily affect
the compliance with the acceptance limits to some extent.
However, the cables used for this purpose will vary, depend-
ing on the application. For operation of the device, the follow-
ing information and conditions should therefore be observed
with regard to interference emission and susceptibility:
1. Data Cables
All connections between measuring instruments or their
interfaces and external devices (printers, computers, etc.)
must always be implemented with adequately shielded
cables. Unless the user manual species a shorter maximum
cable length, the data cable (input/output, signal/control) may
not exceed a total length of 3 m, and may not be located
outdoors. If the device interface allows the connection of
more than one cable, only one may be connected at a time.
Data cables in general should always be double-shielded
cable. For the IEEE bus cable, the double-shielded cable
HZ72 is recommended and can be ordered from ROHDE &
SCHWARZ.
2. Signal Cables
Signal cables used for signal transmission between the
measurement point and the measuring device should gener-
ally be kept as short as possible. Unless a shorter maximum
cable length is specied, the signal cables (input/output,
signal/control) may not exceed a total length of 1 m, and may
not be located outdoors. All signal cables must be shielded
(RG58/U coaxial cable). Care must be taken to achieve a
proper ground connection. For signal generators, double-
shielded coaxial cables (RG223/U, RG214/U) must always be
used.
3. Impact on the Equipment
In the presence of strong high-frequency electric or
magnetic elds, unwanted signal components may enter
the equipment, despite careful setup of the connected
cables and lines. With ROHDE & SCHWARZ measuring
instruments, this will neither be destructive nor lead to the
shutdown of the equipment. As a result of such external
conditions, the displayed values and readings may show
minor deviations with respect to the applicable instrument
specications in individual cases.