Analysis of Volatile Halogenated and
Aromatic Hydrocarbons and
Organophosphorus Pesticides in Water
with a Versatile Agilent 7890B Gas
Chromatography System and an Agilent
7697A Headspace Sampler
Authors
Chunxiao Wang
Agilent Technologies (Shanghai) Co Ltd.
412 YingLun Road
Waigaoqiao Free Trade Zone
Shanghai 200131
P.R. China
Jingqiang Zhang, and ShunNa
Agilent Technologies (China) Co Ltd.
5F, 18F Citic Plaza ShenHong Square
No. 1350 North Sichuan Road
Hongkou District, Shanghai 200080
P.R. China
Application Note
Environmental
Abstract
An Agilent 7890B GC system configured with three detectors: (1) ECD (Electron cap-
ture detector), (2) FID (Flame ionization detector), and (3) FPD (Flame Photometric
Detector) was used for the analysis of volatile halogenated and aromatic hydrocar-
bons and organophosphorus pesticides in water. The headspace sampler (HS) vial
temperature and time for sample equilibration were optimized, allowing simultane-
ous analysis of aromatic and halogenated hydrocarbons with FID and ECD in one HS
injection. The FPD channel with automated liquid sampler (ALS) was used for
organophosphorus pesticides analysis. The repeatability was generally better than
2.0% RSD for each aromatic and halogenated hydrocarbons. The repeatability and
recovery range for organophosphorus pesticides are 1.45–2.74% and 90.4–98.8%,
respectively.
2
Introduction
Volatile halogenated and aromatic hydrocarbons as well as
organophosphorus pesticides have a variety of harmful health
effects. The U.S. Environmental Protection Agency (EPA) esti-
mates that volatile organic compounds are present in
one-fifth of the nation's water supplies. They can enter
ground water from a variety of sources. Benzene, for example,
may enter ground water from gasoline or oil spills on the
ground surface or from leaking underground fuel tanks.
Usually, a large number of samples need to be processed rou-
tinely, and therefore, a high degree of automation is required
with minimal sample prep for cost reduction Static HS/GC is
well suited and widely used for volatile halogenated and
aromatic hydrocarbon determinations. The FPD performs well
for trace organophosphorus pesticide analysis. Therefore, a
7890B GC system with three channels (ECD, FID, and FPD)
forms a versatile system suited for routine analysis of volatile
aromatic and halogenated hydrocarbons and
organophosphorus pesticides in water.
Experimental
Figure 1 shows a 7890B GC system configured with FID, ECD,
and FPD.
Table 1. Typical GC Conditions
Agilent 7697A
Headspace
Sampler
Inlet
Inlet
Splitter
Column 3
Column 1
Column 2
FPD
FID
ECD
Figure 1. Agilent 7890B GC three channel configuration.
(1) FID channel with HS: for aromatic hydrocarbons
(2) ECD channel with HS: for halogenated hydrocarbons
(3) FPD channel with ALS: for organophosphorus pesticides
Volatile aromatic and halogenated organic compounds method
Agilent 7890B GC
Inlet Split/Splitless
Temperature: 150 °C
Split vent: 50 mL/min
Column Column 1: 19091s-413UI, HP-5 MS UI,
30 m × 0.32 mm, 1 µm
Column 2: 123-3234, DB FFAP,
30 m × 0.32 mm, 1 µm
Carrier gas N
2
at 2.5 mL/min constant flow
Oven temperature 40 °C (2 minutes) to 120 °C at 6 °C/min
Detector ECD: Temperature: 320 °C,
Makeup: N
2
at 30 mL/min
FID: Temperature: 200 °C,
H
2
/Air: 40 mL/min/400 mL/min
Agilent 7890B GC7697A Headspace Sampler
20-mL vials
Temperature HS oven: 70 °C
Valve/loop: 70 °C
Transfer line: 100 °C
Time Vial equilibration
time: 40 minutes
Inject time: 0.5 minutes
Vial sampling
parameters
Vial pressurize
value: 15 psi
Vial final sampling
pressure: 10 psi
Loop size: 1 mL
Organic phosphorus method
Inlet Split/Splitless
Temperature: 230 °C
Splitless mode, 30 mL/purge flow at 0.75 minutes
Sample size 1 µL
Column DB-1701, 30 m × 0.25 mm, 1 µm
Carrier gas N
2
at 1 mL/min constant flow
Oven temperature 100 °C to 170 °C at 25 °C/min to 210 °C (1 minute)
at 15 °C/min to 220 °C at 10 °C/min to 240 °C
(5 minutes) at 15 °C/min
Detector FPD Plus
Temperature: 270 °C
Emission block: 150 °C