Self-Calibration with the SDS Series Oscilloscopes
October 25, 2017
Over time and temperature, the components in a test instrument can change and adversely alter the
performance and measurement accuracy. Periodically running the self-calibration routine can help ensure
top performance.
You can run the self-calibration routine at any time, but it is recommended if the operating temperature
changes by more than 5 °C.
1. Connect all probes, generators, or signal sources from the oscilloscope
2. Press the Utility button (On some SDS1000 scopes, the Self Cal softkey may be located on another
menu page. Simply scroll through the Utility menu pages until you find it)
3. Choose “Do Self Cal” and follow the on-screen prompts.