Power Supplies
1601 N. Clancy Ct. • Visalia, CA 93291 • (559) 651-2222 • FAX (559) 651-0188 • sales@tri-mag.com • tri-mag.com 1 of 2
DT080AF Series | ITE & Medical Safety
80W/100W Peak
Level VI energy compliant
No load input power < 0.15 W
Average efciency > 89%
Eco-friendly design
UL/CSA/EN 60950-1, 2
nd
edition
ANSI/AMMI/CSA/EN 60601-1, 3
rd
edition
10 year warranty
The DT080AF Series is a 72W external power supply (EPS) designed for both medical and ITE applications. The eco-friendly
design is compliant with the new DOE Energy Conservation Standards for External Power Supplies. With a 100W peak power
rating, operating efciencies > 89% and no-load input power < 0.15W, the DT080AF Series is ideal for ITE and medical equip-
ment designs, including MOOP and 2xMOPP, that are compliant with 2016 conservation standards.
Specifications
Description
GREEN POWER
Input
Input Voltage 85 VAC to 264 VAC
Input Frequency 47 Hz to 63 Hz
Inrush Current < 40/80A at 115/230 VAC, cold start, 25ºC
Input Protection Internal T3.15A / 250 VAC fuse in line
No Load Input Power < 0.15W
Input Current 2A
rms
max/115 VAC, 1 A
rms
max/230VAC
Output
Output Voltage See tables on page 2
Initial Set Accuracy See tables on page 2
Minimum Load No minimum load required
Start Up Rise Time 2 ms typical
Hold Up Time 18 ms typical
Line Regulation ±0.5% typical
Load Regulation ±3.0% typical
Ripple & Noise < 1% pk-pk typical, 20MHz Bandwidth
Over-voltage Protection latch off
Over-load Protection auto recovery
Short Circuit Protection auto recovery
Environmental
Operating Temperature -20°C to 60°C derating: 2.5% / ºC > 40°C
Cooling 72W, free air convection
Operating Humidity 10-95% RH, non-condensing
Storage Temperature -40°C to +80°C
Altitude 0 to 5000 m
General
Efciency 89% typical
Energy Saving DOE: EPS efciency Level VI
ErP: EC Code of Conduct Ver. 5
(Tier 2/2016)
Isolation 4000 VAC Input to Output, 2 x MOPP
1500 VAC Input to Ground, 1 x MOPP
1500 VDC Output to Ground, 1 x MOPP
Isolation Resistance 50 MΩ
Switching Frequency 120 kHz typical
MTBF > 200 kHrs to MIL-HDBK-217F at 50°C
EMC & Safety
Safety Approvals: UL/CSA/EN 60950-1, 2nd edition
ANSI/AMMI/CSA/EN 60601-1, 3rd edition
CE Mark
Harmonic Currents EN 61000-3-2 class A
EMI EN 55022/CISPR 22 Class B,
EN 61000-3-3
ESD Immunity EN 61000-4-2, 6kV/contact, 8kV/air
Radiated Immunity EN 61000-4-3, 10V/m with 80% AM
EFT Burst EN 61000-4-4, 2kV
Surge EN 61000-4-5, 1kV/L-L, 2kV/L-G
Conducted Immunity EN 61000-4-6, 10V with 80% AM
Magnetic Fields E61000-4-8, 10A/m
Dips & Interruptions EN 61000-4-11, 30% dips 500ms,
60% dips 200ms, 100% dips 10ms,
100% dips 20ms, 100% dips 5000ms
Warranty
Manufacturer’s Warranty 10 years. Call Tri-Mag or go to
www.Tri-Mag.com for details.
Power Supplies
1601 N. Clancy Ct. • Visalia, CA 93291 • (559) 651-2222 • FAX (559) 651-0188 • sales@tri-mag.com • tri-mag.com 2 of 2
DT080AF Series | ITE & Medical Safety
Output Specifications
Model No.
Output
Rail
Load
Initial
Accuracy
Step Efciency
Avg. Eff.
Min Rated Max Peak @20% Load @50% Load @100% Load
DT080AF-5-3 +12V 0A 6.0A - 8.4A +11.4V~+12.6V 89% 89% 89% 89%
DT080AF-8-3 +15V 0A 5.1A - 7.3A +14.25V~+15.75V 89% 89% 89% 89%
DT080AF-6-3 +24V 0A 3.2A - 4.6A +22.8V~+25.2V 89% 89% 89% 89%
DT080AF-14-3 +48V 0A 1.6A - 2.3A +45.6V~+50.4V 89% 89% 89% 89%
Notes
1. Output Load:
Rated 72W for convection cooling.
2. Peak Load Duration:
100W peak rating for durations up to 5 sec. Ideal for motor-starting/in-rush conditions.
3. Engineering Specication:
Contact Tri-Mag for full engineering specication for the specic part number used in your design application.
4. Standby Power Cosumption with System:
New external power supply (EPS) efciency requirements have been established by the U.S. DOE (EPS) and the European ErP
regulations. The no-load power consumption requirements are < 0.21W and < 0.15W respectively.
5. Audible Noise:
For the DT080AF-x-3 energy saving series, achieving level VI (<0.15W) standby power consumption is accomplished through burst
mode operation of the controller. The burst operation frequency is dependent on load conditions and is approx. 114Hz, within the
audible frequency range.
6. Step Efciency and Average Efciency:
Test conditions in step efciency are referred to 3.2.21PS (Internal Power Supply) of ENERGY STAR program requirements for
computers. ENERGY STAR required for efciency @ 20%, 50%, 100% load is 89%, 89%, 89%; average efciency is the average
of step efciency.
7. Model Ordering Table:
Safety/Application Energy Saving
ITE & Medical DT080AF-x-3
2014-10
Mechanical Specifications
Notes
1. Dimensions shown in mm (inch). Tolerance: ±1mm (Excluding cables).
2. Size: 72.0 x 145.0 x 42.0 (mm)
2.83 x 5.70 x 1.65 (inches)
Net weight: 517 g approx. / unit
3. Input Socket: 4. Output Connectors:
IEC320 C14 (Class I)
+12 to +24Vdc Models
+48Vdc Model, DT080AF-14-3
5. Cable length is 1.5m approx.
6. Power ON indication LED is on top of Box. Box Color is Black.
7. RoHS compliant