Data Sheet 202943 Page 1/9
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
E-mail: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow - Essex CM20 2DY, UK
Phone: +44 1279 63 55 33
Fax: +44 1279 63 52 62
E-mail: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard
Canastota, NY 13032, USA
Phone: 315-697-JUMO
1-800-554-JUMO
Fax: 315-697-5867
E-mail: info@jumo.us
Internet: www.jumo.us
Brief description
The sensor detects the electrolytic conductivity of a process liquid. It uses the inductive principle of
measurement. Because it uses the inductive method of measurement rather than the conductive
2-pin or 4-pin measurement method, the sensor is virtually maintenance-free. The measuring accu-
racy is practically unaffected by accumulations and grease or oil film on the surface of the sensor.
A (Pt1000) temperature sensor acquires the process temperature at the same time. The temperature
sensor can either be exposed in a stainless steel pocket (ultra-fast response) or be located right in-
side the plastic body (with PVDF, only the internal option is possible). Internal installation is advisable
if the sample media could chemically attack the stainless steel pocket.
The standard sensor body material is polypropylene (PP). PVDF is also available as an alternative.
A vast number of process connections are available to ensure system flexibility, even to provide
spare equipment for older devices.
The JUMO ecoLine Ci sensor is designed for connection to the following transmitters: JUMO
AQUIS 500 Ci (as per data sheet 202566), JUMO AQUIS touch S/P (as per data sheet 202580/
202581), and JUMO digiLine Ci ST10 (as per data sheet 202760).
Typical areas of application:
The sensor is primarily designed for use in water engineering systems. Typical areas of application
include general water engineering (drinking water and wastewater), air-conditioning and cooling
plants, dilution monitoring in cooling towers, flushing tanks, carwashes, sea water desalination (inlet),
monitoring swimming pool water, etc. There is an immersion variant that can be used in open chan-
nels and containers.
Product advantages:
Practically maintenance-free conductivity measurement
Materials are FDA/food-use approved
A variety of process connection variants
A fast-response temperature sensor
A compact, proven sensor
Type 202943/10-...
Type 202943/10-... with
T-piece made from PVC
JUMO ecoLine Ci
Inductive conductivity and temperature sensor for
general water engineering
V3.00/EN/00556192/2022-02-16 20294300T10Z001K000
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
E-mail: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow - Essex CM20 2DY, UK
Phone: +44 1279 63 55 33
Fax: +44 1279 63 52 62
E-mail: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard
Canastota, NY 13032, USA
Phone: 315-697-JUMO
1-800-554-JUMO
Fax: 315-697-5867
E-mail: info@jumo.us
Internet: www.jumo.us
Data Sheet 202943 Page 2/9
V3.00/EN/00556192/2022-02-16 20294300T10Z001K000
Technical data
Type 202943/10-... 202943/20-... 202943/30-...
Conductivity measurement
principle
Inductive
Conductivity meas. range
a
a
Usual application range from approx. 100 µS/cm.
0 ... 1000 µS/cm to 0 ... 2000 mS/cm (depending on connected transmitter)
Conductivity accuracy
for measuring range:
0-1mS/cm
0-10mS/cm
0-50mS/cm
0-100mS/cm
0 - 1000 mS/cm
0 - 2000 mS/cm
1%
0.5 %
0.5 %
0.5 %
1%
1%
Cell constant k = 6.25 1/cm k = 4.65 1/cm
Temperature sensor Pt1000, Class A
t
90
temperature
b
b
According to EN 60751
6s 2min 10 min
Permissible ambient temperature -10 to +60 °C
Permissible storage temperature -20 to +75 °C
Enclosure protection
c
c
According to EN 60529
IP68
Permissible medium temperature
d
In operation
Momentary
with process connections 168
and 706 (immersion fitting)
d
Note: Temperature, pressure and sample medium affect the life of the cell!
-10 to +80 °C
+100 °C ( 15 min)
-10 to +60 °C (pressureless)
-10 to +100 °C
+110 °C ( 15 min)
-10 to +60 °C (pressureless)
Permissible process pressure
c
at +20 °C
at +60 °C
at +80 °C
at -10 to +100 °C
10 bar
6bar
0bar
min. -0.1 bar
10 bar
6bar
4bar
min. -0.1 bar
Sensor material
in contact with medium
not in contact with medium
dependent on design: stainless
steel 1.4301 (304), stainless
steel 1.4571 (316ti), PP, EPDM
dependent on design: stainless
steel 1.4301 (304), PA6, PUR,
FPM, PBT/PA, CR/NBR, CuZn
PP
dependent on design: stainless
steel 1.4301 (304), PA6, PUR,
FPM, PBT/PA, CR/NBR, CuZn
PVDF
dependent on design: stainless
steel 1.4408, PA6, PUR, PBT/PA,
CR/NBR, CuZn
Process connection see Order details / Dimensions
Electrical connection
Connection circuit
Socket
Socket material
Cable material
Cable lengths
Permissible temperature
JUMO ecoLine Ci type inductive conductivity sensors are suitable for connection to JUMO AQUIS 500
Ci type inductive conductivity transmitters/controllers!
non-detachable cable
M12 socket, 8-pin
CuZn, PA6.6 GF30, PUR
outer sleeve: PUR
see Order details
-20 to +75 °C
Approvals/marks of conformity Materials in contact with the medium are approved for food use, are physiologically safe and are listed
by the FDA.
0
2
4
6
8
10
-10
0
10
20
30
40
50
60
70
80
90
100
T
e
mperature
[°C]
Pressure [bar]
0
2
4
6
8
10
-10
0
10
20
30
40
50
60
70
80
90
100
T
e
mperature
[°C]