Calibration
WIKA data sheet CT 15.01
Page 1 of 17
WIKA data sheet CT 15.01
05/2022
Data sheets showing similar products and accessories:
Test pumps, hydraulic; model CPP series; see data sheet CT 91.05
Test pump, pneumatic; model CPP30; see data sheet CT 91.06
Calibration software; WIKA-Cal; see data sheet CT 95.10
Pressure calibrator
Standard version
Model CPH6000
Pressure calibrator, model CPH6000
Above: With rmly mounted reference pressure sensor
Below: Reference pressure sensor via sensor
connection cable
Description
The model CPH6000 is operated with a calibration assistant
that automatically ensures complete documentation of
all measuring results. The pressure calibrator enables
completely paperless calibration on-site via a pressure
sensor. The hand-held recognises the various reference
sensors and relieves the user from having to congure them
via the menu.
The reference pressure sensor model CPT6000 is available
for the respective suitable application solution. The CPT6000
features accuracy down to 0.025 % and measuring ranges
up to 6,000 bar [87,000 psi]. It can be replaced on the
CPH6000 quickly and without using any tools. In addition to
being xed to the instrument, the CPT6000 can, optionally,
be used via an approx. 1.1 m [3.6 ft] long sensor connection
cable.
The pressure calibrator can be a complete measuring
chain consisting of the various individual components or a
complete test and service case (see page 15).
The matching test pump is optionally available, depending on
the case set.
For the evaluation and documentation of the calibration
data stored in the CPH6000, WIKA-Cal calibration software
is available. Using this software, the data is automatically
transferred into a printable calibration certicate. For data
transfer, an RS-232 and a USB interface are available.
for further approvals,
see page 11
Applications
Calibration service companies and service industry
Maintenance facilities
Measurement and control laboratories
Quality assurance
Special features
Digital indicator with easily interchangeable reference
pressure sensors model CPT6000 (sensor can also be
attached externally)
Measuring ranges of sensor: from -1 ... 6,000 bar
[-14.5 ... 87,000 psi]
Accuracy: up to 0.025 % (incl. calibration certicate)
Easy handling for calibration and pressure switch test
function with calibration assistant
Software compatible with WIKA-Cal
WIKA data sheet CT 15.01 ∙ 05/2022 Page 2 of 17
Functionalities
Features
15 pressure units + 1 programmable unit
Programmable functions: Min./Max./Tare/Filter/Alarm/
Altitude correction
The display of the test item measured value is also
possible as an original electrical signal (mA/V)
Applications
Measurement of operating/process pressures
Comparative measurements with test items (power supply
and display for the test item through the CPH6000)
Min. and Max. memory (e.g. for leak testing)
Alarm function for safety testing
Operating mode: CALIBRATION
Features
Calibration assistant
Programmable altitude correction
Recordable calibration temperature
Applications
On-site calibration of pressure sensor and pressure
measuring instruments (without PC)
A calibration assistant guides you easily through the
calibration (in line with DAkkS). With this, the data sets,
including date and time, are recorded within the CPH6000.
Prior to calibration, the calibration steps are entered
directly on the instrument or downloaded via WIKA-Cal
software.
Available PC software
WIKA-Cal calibration software for creating calibration
certicates
mA
123456
ID Nr:
01
P-01:
bar
0.000
Ist:
bar
0.000
Soll:
%FS0.07
bar
0.004
Abw.:
bar
0.004
6.000
0.000
Prüfling
bar
0.000
10.000
0.000
Referenz
13:41
KALIBRIEREN
A) Record calibration data with customer-specic test item
USB/RS-232
Print
Calibration certificate
Kalibrierzeugnis
Certificate no. 00000039
Zeugnis-Nr.
Page 1 / 3
Seite
Demo laboratory
Demo street 123
12345 Demo city
Demo country
Tel.: 01234/56789
Fax: 01234/56789
E-Mail: demo@demo.com
www.demo.com
General
Allgemein
Device under test (DUT)
Kalibriergegenstand (KG)
Calibration procedure
Kalibrierverfahren
Calibration certificate
Kalibrierschein
Customer Demo customer
Kunde
Demo street
DE-12345 Demo city
Customer order number
Kundenbestellnummer
Order date
Bestelldatum
Date of calibration 15.02.2021
Datum der Kalibrierung
Operator Demo User
Bearbeiter
Electrical pressure gauge
Elektrisches Druckmessgerät
Model D-10
Typ
Serial number demo12345
Seriennummer
Scale range (0 ... 10) bar g
Anzeigebereich
Accuracy 0.05 % FS
Genauigkeit
Resolution (scale division) 0.0001 bar
Auflösung (Skalenwert)
TagNo demo tag
Messstellen-Nr.
Calibration is carried out in accordance with the following standard: IEC 61298
Die Kalibrierung erfolgt auf der Grundlage der folgenden Norm: IEC 61298
B) Record calibration data on the PC and print the calibration
certicate
In the setup menu there are 3 operating modes to choose
from:
MEASURING, CALIBRATION and SWITCH TEST. In the
CALIBRATION and SWITCH TEST modes, a menu assistant
supports the user during each operation and records, for
example, the calibration data from several calibrations or
automatically calculates the switch hysteresis.
To power the test item and to read its measuring signals,
there are electrical inputs and outputs which are protected
from adverse conditions in the eld by captive protective
caps.
Operating mode: MEASURING
%FS
0.10
bar
0.006
Abw.:
0.006
bar
6.000
0.000
Prüfling
bar
0.000
10.000
0.000
Referenz
13:28
MESSEN
Example: CPH6000 with rmly mounted reference pressure
sensor, hand spindle pump and customer-specic test item