1
®
AN143.0
CAUTION: These devices are sensitive to electrostatic discharge; follow proper IC Handling Procedures.
1-888-INTERSIL or 1-888-468-3774
| Intersil (and design) is a registered trademark of Intersil Americas Inc.
Copyright Intersil Americas Inc. 2005. All Rights Reserved
All other trademarks mentioned are the property of their respective owners.
X3100 Voltage Measurement Accuracy
Description
This test measures the accuracy of the celll voltage
measurement circuits of the X3100. In the test, a known
voltage is applied across each cell input of the X3100. The
analog output voltage of the X3100 is measured for each
cell. This value is compared to the input voltage. The
collected data includes the values from three devices from
three different lots. The data also includes tests at 0°C, 25°C
and 50°C.
Test Setup
For this test, Vcc was held at 16V to prevent the device
entering the sleep mode when testing cell voltages below
2V. In the test 4V was applied to all cells, except the cell
being measured. The voltage was applied at 10mV intervals
over four ranges. These input ranges were 1.5 to 1.7V, 2.5 to
2.7V, 3.5 to 3.7V and 4.0 to 4.2V.
For each cell and each applied voltage the output was
recorded. This involves the changing of the analog
multiplexer setting in the X3100. In gathering the data with
this test, the output was sample. This procedure was used to
reduce the effect of external noise on the test environment.
Results
The data for this test is available in raw form in an Excel
spreadsheet. This spreadsheet includes data for each
device, data for each lot, and data for all devices. Each set of
data includes results for the three temperature settings and
the four gain settings. The data was copied to other files for
the purposes of generating graphs.
The result summaries show the output error as a
percentage. This error is provided for each of 3 lots as well
as the combined data. Data is averaged, and standard
deviation computed, from samples taken at three
temperatures 0°C, 25°C and 50°C.
On the following charts, absolute error is the Output minus
the Input, while the percent error is calculated as follows:
Overall Performance
The following set of graphs shows the performance of the
X3100 over temperature and across three lots of three
devices. The graphs will show the basic input/output curve,
raw error, and error after correction for offset and gain.
The below data shows that the X3100 output is accurate,
with 3 standard deviations, to better than 2.5% error across
all measured voltages. It also shows that over the key range
of 2.5 to 4.2V, accuracies of better than 1% can be expected.
These accuracies are not dependent on any calibration
factors.
X3100
+
+
+
+
16V
AO
VCELL1
VCELL2
VCELL3
VCELL4
VSS
Error
Output Input
Input
-----------------------------------------=
Application Note May 9, 2005
FIGURE 1. CELL VOLTAGE MEASUREMENT
2
AN143.0
May 9, 2005
HP6626
Power Supply
X3100
+
-
+
-
+
-
+
-
VCELL1
VCELL2
VCELL3
VCELL4
VSS
Keithley 236
Source Measurement Unit
HP6626
Power Supply
Keithley 2002
DMM
VCC
VRGP
VRGC
VRGO
Keithley 2000
DMM
AO
PC
SPI Serial
Analog Select
VCS1 VCS2
20.37
Ohm
FIGURE 1. CHARACTERIZATION TEST SETUP
Application Note 143