Model 2606B
System SourceMeter Instrument Specifications
Specifications are subject to change without notice
SPEC-2606B Rev. C February 2021
*PSPEC-2606BC*
1 of 9
Keithley Instruments
28775 Aurora Road
Cleveland, Ohio 44139
1-800-833-9200
tek.com/keithley
SPECIFICATION CONDITIONS
This document contains specifications and supplemental information for the Model 2606B System
SourceMeter
®
instrument. Specifications are the standards against which the 2606B instruments are
tested. Upon leaving the factory, the 2606B instruments meet these specifications. Supplemental and
typical values are nonwarranted, apply at 23 °C, and are provided solely as useful information.
Specifications are for individual modules.
Source and measurement accuracies are specified at the 2606B terminals under these conditions:
18 °C to 28 °C, < 70 percent relative humidity
After a two-hour warm-up period
Speed normal (1 NPLC)
A/D autozero enabled
Remote sense operation or properly zeroed local operation
Calibration period: One year
DC POWER SPECIFICATIONS
Voltage
Current
Maximum output
power and source
limits
1,2
20.2 W maximum
20.2 V at 1.0A, -20.2 V at -1.0 A
6.06 V at 3.0 A, -6.06 V at -3.0A
20.2 W maximum
1.01 A at 20 V, -1.01 A at -20 V
3.03 A at 6 V, -3.03 A at -6 V
1
For additional power derating information for various load and temperature conditions, refer to “Operating Boundaries” in the Series 2606B
Reference Manual.
2
The System SourceMeter
®
will allow quadrant II and quadrant IV operation, however, it is intended for low power or short transient behavior.
Operation in quadrant II and quadrant IV could result in an overtemperature error.
Model 2606B System SourceMeter
®
Instrument Specifications
Specifications are subject to change without notice
2 of 9 SPEC-2606B Rev. C February 2021
VOLTAGE ACCURACY SPECIFICATIONS
3
,
4
Range
Source
Programming
resolution
Accuracy
± (% reading + volts)
Typical noise
(peak to peak)
0.1 Hz to 10 Hz
Display
resolution
Accuracy
5
± (% reading + volts)
100 mV
5 µV
0.02% + 250 µV
20 µV
100 nV
0.015% + 150 µV
1 V
50 µV
0.02% + 400 µV
50 µV
1 µV
0.015% + 200 µV
6 V
50 µV
0.02% + 1.8 mV
100 µV
1 µV
0.015% + 1 mV
20 V
500 µV
0.02% + 12 mV
500 µV
10 µV
0.015% + 8 mV
CURRENT ACCURACY SPECIFICATIONS
3
Range
Source
Measure
Programming
resolution
Accuracy
± (% reading + amperes)
Typical noise
(peak to peak)
0.1 Hz to 10 Hz
Display
resolution
Accuracy
5
± (% reading + amperes)
100 nA
2 pA
0.1% + 100 pA
5 pA
100 fA
0.08% + 100 pA
1 µA
20 pA
0.03% + 800 pA
25 pA
1 pA
0.025% + 500 pA
10 µA
200 pA
0.03% + 5 nA
60 pA
10 pA
0.025% + 1.5 nA
100 µA
2 nA
0.03% + 60 nA
3 nA
100 pA
0.02% + 25 nA
1 mA
20 nA
0.03% + 300 nA
6 nA
1 nA
0.02% + 200 nA
10 mA
200 nA
0.03% + 6 µA
200 nA
10 nA
0.02% + 2.5 µA
100 mA
2 µA
0.03% + 30 µA
600 nA
100 nA
0.02% + 20 µA
1 A
20 µA
0.05% + 1.8 mA
70 µA
1 µA
0.03% + 1.5 mA
3 A
20 µA
0.06% + 4 mA
150 µA
1 µA
0.05% + 3.5 mA
3
For temperatures 0 °C to 18 °C and 28 °C to 50 °C, accuracy is degraded by ±(0.15 × accuracy specification)/°C.
4
Add 50 µV to source accuracy specifications per volt of HI lead drop.
5
Derate accuracy specification for NPLC setting < 1 by increasing the error term. Add appropriate typical percent of reading term for resistive
loads using the table below.
NPLC setting
100 mV range
1 V and 40 V ranges
100 nA range
1 µA to 100 mA ranges
1 A to 3 A ranges
0.1
0.01%
0.01%
0.01%
0.01%
0.01%
0.01
0.08%
0.07%
0.1%
0.05%
0.05%
0.001
0.8%
0.6%
1%
0.5%
1.1%